Medtronic Neuromodulation
7000 Central Ave Ne, Minneapolis, Minnesota, United States
FEI 2182207
FDA has 15 inspections on file for Medtronic Neuromodulation in Minneapolis, Minnesota (FEI 2182207), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection: 2 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Apr 3, 2013. FDA issued a warning letter on (CDRH). FDA's compliance record also lists one other action under this FEI: Injunction, . 3 import refusals are filed under this FEI, the latest on .
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 15, 2009–2023
- 2 OAI · 8 VAI · 5 NAI
- Inspections with 483 observations shown
- 7 of 15
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Aug 9, 2024
- Warning letters
- · CDRH ↗
- Import refusals
- 3 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
15 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Time to report - 10 days
A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for timely and effective identification of events that may be subject to medical device reporting requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design input - documentation
Design input requirements were not adequately documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Procedures not adequately established or maintained
Document control procedures have not been adequately established .
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch MonitoringCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Design validation - production units
The design was not validated using initial production units, lots or batches or their equivalents.
Design input - documentation
Design input requirements were not adequately documented.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- NAINo objectionable conditions that need follow-upDrugs · Postmarket Surv. and EpidemiologyNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Aug 9, 2024. It asked for a Form 483.
Inspection ending May 5, 2023VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Aug 9, 2024FDA-FOIA-2024-7123Form 483· 1 of 5 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 15 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action7 of 7
- 21 CFR 806.10Reports of corrections and removals3 of 7
- 21 CFR 820.30Design controls3 of 7
- 21 CFR 820.198Complaint files3 of 7
- 21 CFR 820.90Nonconforming product2 of 7
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 7
- 21 CFR 820.40[Reserved]1 of 7
- 21 CFR 820.75Process validation1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- InjunctionCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2182207, one row per refused product line.
3 refusals on 2017-08-29 · charged FRNMFGREG on 3 · no FDA sample analysis
FDA firm profile- 2017-08-29
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE80LKK
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsUPS-4651684-6/91/1
- 2017-08-29
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF84LGW
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsUPS-4651684-6/51/1
- 2017-08-29
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)84GZB
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsUPS-4651684-6/41/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2182207 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2182207. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.