GE Medical Systems Information Technologies, Inc.

8200 W Tower Ave, Milwaukee, Wisconsin, United States

FEI 2124823

FDA has 4 inspections on file for GE Medical Systems Information Technologies, Inc. in Milwaukee, Wisconsin (FEI 2124823), from 2010 to 2018. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 5 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below.

2 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.30(f), 21 CFR 820.30(g) and 21 CFR 820.100(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2010–2018
0 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
2 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    2. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.

    3. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    4. Device number, identification

      The total number of devices subject to correction or removal actions was not reported.

    5. Design validation - software validation not performed

      Validation of device software is inadequate and is incomplete.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Notification of extended action

      No amended notification with required information was submitted within 10 working days of initiating an extension of the correction or removal action.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Records of complaint investigation

      Records of complaint investigations do not include required information.

    4. Service reports

      Service reports do not include the required information.

    5. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Complaints

      Complaint files are not adequately maintained.

    2. Submission of Supplement to Individual Report Within One Month

      A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

    3. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    4. Design validation - software validation documentation

      Results of the validation of the device software were not adequately documented.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 1 more citation
    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

  4. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

  • 21 CFR 806.10Reports of corrections and removals3 of 3
  • 21 CFR 820.30Design controls3 of 3
  • 21 CFR 820.100Corrective and preventive action3 of 3
  • 21 CFR 820.198Complaint files2 of 3
  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
  • 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 3
  • 21 CFR 820.200Servicing1 of 3

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2124823, one row per refused product line.

2 refusals from 2017-03-16 to 2018-03-09 · charged FRNMFGREG on 1, REGISTERED on 1 · no FDA sample analysis

FDA firm profile
  • 2018-03-09

    MONITOR, PHYSIOLOGICAL, PATIENT74MHX

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsAEK-4133868-9/11/1

  • 2017-03-16

    ASSEMBLY, TUBE HOUSING, X-RAY, DIAGNOSTIC90ITY

    NO REGISTR 536(a); Failure to file initial report

    Division of Southeast Imports799-9207017-2/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2124823 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2124823. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.