GE Medical Systems Information Technologies, Inc.
8200 W Tower Ave, Milwaukee, Wisconsin, United States
FEI 2124823
FDA has 4 inspections on file for GE Medical Systems Information Technologies, Inc. in Milwaukee, Wisconsin (FEI 2124823), from 2010 to 2018. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 5 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below.
2 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.30(f), 21 CFR 820.30(g) and 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2010–2018
- 0 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 2 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Device number, identification
The total number of devices subject to correction or removal actions was not reported.
Design validation - software validation not performed
Validation of device software is inadequate and is incomplete.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Notification of extended action
No amended notification with required information was submitted within 10 working days of initiating an extension of the correction or removal action.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Records of complaint investigation
Records of complaint investigations do not include required information.
Service reports
Service reports do not include the required information.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complaints
Complaint files are not adequately maintained.
Submission of Supplement to Individual Report Within One Month
A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Design validation - software validation documentation
Results of the validation of the device software were not adequately documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 1 more citation
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 806.10Reports of corrections and removals3 of 3
- 21 CFR 820.30Design controls3 of 3
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 3
- 21 CFR 820.200Servicing1 of 3
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2124823, one row per refused product line.
2 refusals from 2017-03-16 to 2018-03-09 · charged FRNMFGREG on 1, REGISTERED on 1 · no FDA sample analysis
FDA firm profile- 2018-03-09
MONITOR, PHYSIOLOGICAL, PATIENT74MHX
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-4133868-9/11/1
- 2017-03-16
ASSEMBLY, TUBE HOUSING, X-RAY, DIAGNOSTIC90ITY
NO REGISTR 536(a); Failure to file initial report
Division of Southeast Imports799-9207017-2/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2124823 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2124823. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.