Electromedical Products International, Inc.
2201 Garrett Morris Pkwy, Mineral Wells, Texas, United States
FEI 2020648
FDA has 4 inspections on file for Electromedical Products International, Inc. in Mineral Wells, Texas (FEI 2020648), from 2009 to 2018. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Feb 1, 2013. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2009–2018
- 1 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of procedures, or not maintained
Document control procedures have not been maintained.
Approval, inadequate purchasing data
Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier oversight
The type and extent of control to be exercised over the product and suppliers was not clearly defined.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nature and details of complaint
Records of complaint investigations do not include the nature and details of the complaint.
Management review accomplishment
Management reviews do not ensure that the quality system satisfies the requirements of part 820 and your quality policy and objectives.
Specific non-conforming product procedures
Procedures for addressing the identification, documentation, evaluation, and investigation of nonconforming product were not defined, documented, and complete.
Lack of or inadequate DHR procedures
Procedures have not been established to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
Uniform and timely processing
Complaint handling procedures have not been defined to ensure that all complaints are processed in a uniform and timely manner.
Show 4 more citations
Documentation
Corrective and preventive action activities have not been documented, including the actions needed to correct or prevent recurrence of nonconforming product and other quality problems and submission of information on quality problems and corrective and preventive actions for management review.
DMR QA procedures and specifications
The device master record does not include or refer to the location of all quality assurance procedures and specifications.
Records accessibility
Investigated complaints and records of investigation were not accessible to the manufacturing establishment.
DMR - not or inadequately maintained
The device master record does not include or refer to the location of installation, maintenance, and servicing procedures and methods.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.90Nonconforming product1 of 3
- 21 CFR 820.181Device master record1 of 3
- 21 CFR 820.184Device history record1 of 3
Show all 9
- 21 CFR 820.198Complaint files1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2020648 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2020648. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.