Amrex Zetron Inc
7034 Jackson St, Paramount, California, United States
FEI 2011115
FDA has 4 inspections on file for Amrex Zetron Inc in Paramount, California (FEI 2011115), from 2008 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 5 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 801, the device labeling rules, and 1 under 21 CFR Part 830, the unique device identification rules. They are listed below.
The most recent OAI inspection ended on Nov 3, 2008. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.22, 21 CFR 820.184 and 21 CFR 820.200(c).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2008–2024
- 1 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Service reports/MDRs/complaints
Service reports that represent MDR reportable events were not automatically considered complaints and processed in accordance with the requirements of 21 CFR 820.198.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Service reports/MDRs/complaints
Service reports that represent MDR reportable events were not automatically considered complaints and processed in accordance with the requirements of 21 CFR 820.198.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audit/Reaudit - conducted
Quality audits were not conducted to verify that the quality system is effective in fulfilling your quality system objectives.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure that equipment is routinely calibrated were not established and implemented.
Analyzing service report
Service reports were not analyzed following appropriate statistical methods.
Quality System Procedures
Quality system procedures were not established and implemented.
Management responsibility
Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.
Show 7 more citations
Lack of Written MDR Procedures
Written MDR procedures have not been developed and implemented.
Management review - defined interval, sufficient frequency
Management reviews were not conducted at defined intervals.
Servicing - Lack of or inadequate procedures
Servicing procedures have not been established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Results of investigation
Records of complaint investigations do not include the results of the investigation.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy has been fully implemented and maintained at all levels of the organization.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit3 of 3
- 21 CFR 820.200Servicing3 of 3
- 21 CFR 820.184Device history record2 of 3
- 21 CFR 801.20Label to bear a unique device identifier1 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
Show all 10
- 21 CFR 820.198Complaint files1 of 3
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2011115 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2011115. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.