Global Medical Holdings LLC
26378 289th Pl, Adel, Iowa, United States
FEI 1931654
FDA has 4 inspections on file for Global Medical Holdings LLC in Adel, Iowa (FEI 1931654), from 2009 to 2014. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 9 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Sep 6, 2012. No warning letter, court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2009–2014
- 1 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complaints
Complaint files are not adequately maintained.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.
Show 4 more citations
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
DMR - not or inadequately maintained
A device master record has not been maintained.
DHR documentation of label release {see also 820.184}
The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
DHR documentation of label release {see also 820.184}
The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Records of MDR Investigation
Investigation records of MDR reportable complaints do not include required information.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 820.120Device labeling2 of 3
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.86Acceptance status1 of 3
- 21 CFR 820.181Device master record1 of 3
- 21 CFR 820.184Device history record1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1931654 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1931654. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.