UCH-MHS

1400 E Boulder St, Colorado Springs, Colorado, United States

FEI 1770154

FDA has 6 inspections on file for UCH-MHS in Colorado Springs, Colorado (FEI 1770154), from 2010 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection, all under 21 CFR Part 606, the manufacturing rules for blood and blood components. They are listed below.

The most recent OAI inspection ended on Sep 3, 2010. No warning letter, court action or import refusal is filed under this FEI.

Eight regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2010–2024
1 OAI · 3 VAI · 2 NAI
Inspections with 483 observations shown
4 of 6
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the processing of blood and blood components for allogeneic transfusion were not always followed.

    2. Concurrent documentation

      Records are not concurrently maintained with the performance of each significant step in the processing of each unit of blood and blood components so that all steps can be clearly traced.

  2. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  3. NAINo objectionable conditions that need follow-upBiologics · Blood and Blood ProductsNone
  4. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Biological product deviation report

      Failure to submit a biological product deviation report.

    2. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the compatibility testing, storage, and distribution of blood and blood components for homologous transfusion are not always followed.

    3. Thorough investigations

      Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy.

    4. Required records

      Failure to maintain storage and distribution, quality control, and general records.

    5. Concurrent documentation

      Records are not concurrently maintained with the performance of each significant step in the storage and distribution of each unit of blood and blood components so that all steps can be clearly traced.

    Show 3 more citations
    1. Written SOPs available for use by personnel

      Failure to make available written procedures for use by personnel in the areas where the procedures are performed.

    2. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.

    3. Required records - transfusion reaction and complaints

      Failure to maintain records of transfusion reaction reports and complaints, including investigation and follow up.

  5. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products28 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Explanation of hazards

      Failure to explain to the prospective Source Plasma donor the hazards of the plasmapheresis procedure.

    2. Obtaining consent in an appropriate manner

      Failure to present the hazards and risks of plasmapheresis in such a manner that intelligent and informed consent can be given by the prospective Source Plasma donor.

    3. Notification w/in 8 weeks

      Failure to make reasonable attempts to notify the donor within 8 weeks after determining that the donor is deferred or determined not to be suitable for donation.

    4. Donor suitability by means of medical history

      Failure to determine donor suitability on the day of collection by means of test for hemoglobin level.

    5. Corrective action

      Failure to take immediate corrective action and maintain a record of corrective action when the quality control testing for platelets does not meet the prescribed requirements.

    Show 23 more citations
    1. Documentation

      Failure to document that you have successfully notified a deferred donor and document that you have made reasonable attempts to notify a deferred donor.

    2. Notification

      Failure to make reasonable attempts to notify a donor who has been deferred based on the results of tests for evidence of communicable disease agent(s) and determined not be to suitable as a donor based on suitability criteria.

    3. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not fully followed.

    4. Required records

      Failure to maintain storage and distribution records.

    5. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection, processing, storage, and distribution of blood and blood components for homologous transfusion and autologous transfusion are not always maintained and followed.

    6. Lab controls established, including changes

      The establishment of sampling plans including any changes thereto, are not drafted by the appropriate organizational unit and reviewed and approved by the quality control unit.

    7. Supervision of qualified physician

      Performance of plateletpheresis not under the supervision of a qualified licensed physician who was aware of the health status of the donor.

    8. Accurate measurement of quantity of blood

      The standard operating procedure fails to include a written description of the blood collection procedure, including in-process precautions taken to measure accurately the quantity of blood removed from the donor.

    9. Investigations of discrepancies, failures

      There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.

    10. Written certification

      Performance of plateletpheresis without a qualified licensed physician certifying in writing that the donor's health permitted plateletpheresis.

    11. Adverse Reaction- Reports of Investigations

      Written reports of investigations of adverse reactions, including conclusions and follow up, are not prepared and maintained.

    12. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate sampling plans designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity.

    13. Donor deferral record - each location

      A record is not available from which unsuitable (deferred) donors may be identified so that products from such individuals will not be distributed.

    14. Person performing, test results, interpretation

      Records fail to identify the person performing the work, include dates of the various entries, and be as detailed as necessary so as to provide a complete history of the work performed.

    15. Arm preparation

      The phlebotomy site is not prepared by a method that gives maximum assurance of a sterile container of Whole Blood.

    16. Donor suitability procedures not followed

      Failure to follow standard procedures and methods for determining the suitability of a donor as a source of blood.

    17. Lack of quality control unit

      There is no quality control unit.

    18. GMP Training Frequency

      GMP training is not conducted on a continuing basis and with sufficient frequency to assure that employees remain familiar with CGMP requirements applicable to them.

    19. Storage temperatures after collection

      After collection, blood is not transported from the donor clinic to the processing laboratory in temporary storage to cool the blood continuously toward a range between 1 and 6 degrees Celsius.

    20. Concurrent documentation

      Records are not concurrently maintained with the performance of each significant step in the processing of each unit of blood and blood components so that all steps can be clearly traced.

    21. General requirements - inspection

      Failure to inspect Red Blood Cells immediately after separation of the plasma, periodically during storage, and at the time of issue.

    22. Thorough investigations

      Failure to perform a thorough investigation and make a record of the conclusions and follow-up of an unexplained discrepancy and a failure of a lot or unit to meet any of its specifications.

    23. Controlling storage temperatures

      The standard operating procedure fails to include a written description of the storage temperatures and methods of controlling storage temperatures for all blood products and reagents

  6. OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Blood and Blood Products16 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Accurate measurement of quantity of blood

      The standard operating procedure fails to include a written description of the blood collection procedure, including in-process precautions taken to measure accurately the quantity of blood removed from the donor.

    2. Notification of transfusion recipients

      Failure to notify the attending physician of a recipient of a lookback unit.

    3. Who must report - manufacturer

      Failure to submit biological deviation reports.

    4. Results of testing

      Failure to notify the donor of the results of tests for evidence of infection due to communicable disease agents that were a basis for deferral and supplemental tests.

    5. Establish, maintain and follow manufacturing SOPs

      Written standard operating procedures including all steps to be followed in the collection and processing of blood and blood components for homologous transfusion and further manufacturing purposes are not always followed.

    Show 11 more citations
    1. Thorough investigations

      Failure to perform a thorough investigation of an unexplained discrepancy and a failure of a lot or unit to meet any of its specifications.

    2. Donor Deferral - reentry

      The method or process used to requalify deferred donors was not found acceptable for such purposes by FDA.

    3. Required records

      Failure to maintain donor, processing, storage and distribution, compatibility testing, quality control, and general records.

    4. Storage temps./agitation

      Failure to store platelets immediately after resuspension at 20 to 24 degrees Celsius with continuous gentle agitation.

    5. Controlling storage temperatures

      The standard operating procedure fails to include a written description of the storage temperatures and methods of controlling storage temperatures for all blood products and reagents

    6. Concurrent documentation

      Records are not concurrently maintained with the performance of each significant step in the collection of each unit of blood and blood components so that all steps can be clearly traced.

    7. SOP compliance

      Failure of the Standard Operating Procedure to comply with additional standards in 21 CFR 640.

    8. Procedure for HIV or HCV lookback

      There is no written procedure to look at prior donations of Whole Blood, blood components, Source Plasma, and Source Leukocytes, when a donor of blood subsequently tests repeatedly reactive for antibody to human immunodeficiency virus, or otherwise is determined to be unsuitable when tested in accordance with the applicable regulations.

    9. Following manufacturer's instructions

      Failure to use supplies and reagents in a manner consistent with instructions provided by the manufacturer.

    10. Qualifications of responsible personnel

      The personnel responsible for the collection of blood or blood components are not adequate in training and experience, including professional training as necessary to assure competent performance of their assigned functions, and to ensure that the final product has the safety, purity, potency, identity and effectiveness it purports or is represented to possess.

    11. Corrective action

      Failure to take immediate corrective action when the quality control testing for platelets does not meet the prescribed requirements.

Regulations FDA cited here

The 20 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.

Show all 20

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1770154 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1770154. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.