Skip to content

Ultradent Products, Inc.

505 W 10200 S, South Jordan, Utah, United States

FEI 1718912

FDA has 7 inspections on file for Ultradent Products, Inc. in South Jordan, Utah (FEI 1718912), from 2010 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below. Three of the four FOIA requests for this site have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Nov 8, 2013. FDA issued a warning letter on (CDRH). 9 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.

Four regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2010–2025
1 OAI · 3 VAI · 3 NAI
Inspections with 483 observations shown
6 of 7
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
4 requests, latest May 14, 2026
Import refusals
9 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Written procedures not established/followed

      Written procedures are not followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product.

    2. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate specifications and test procedures designed to assure that components and drug products conform to appropriate standards of identity, strength, quality and purity.

    3. No written record of investigation

      Written records are not made of investigations into unexplained discrepancies.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Test methods

      The accuracy and reproducibility of test methods have not been established.

    2. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate specifications designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity.

    3. Following/documenting laboratory controls

      Established test procedures and laboratory control mechanisms are not followed and documented at the time of performance.

  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Buildings not maintained in good state of repair

      Buildings used in the manufacturing and processing of a drug product are not maintained in a good state of repair.

  5. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Written record of investigation incomplete

      Written records of investigations into unexplained discrepancies do not always include the conclusions and follow-up.

    2. Control procedures to monitor and validate performance

      Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.

    3. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Design verification - documentation

      The design verification results, including method(s), the date, and the individual(s) performing the verification, were not documented in the design history file.

    4. Design input - review and approval

      Design input requirements were not reviewed and approved by designated individual(s).

    5. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    Show 3 more citations
    1. Design input - Lack of or inadequate procedures

      Procedures for design input have not been established.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not where necessary.

    3. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

  7. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Complete files maintained

      Complete complaint files are not maintained.

    2. Design validation- Lack of or inadequate procedures

      Procedures for validating the device design were not implemented.

    3. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    4. Written record of investigation incomplete

      Written records of investigations into the failure of a batch or any of its components to meet specifications do not always include the conclusions and follow-up.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Four requests to FDA under the Freedom of Information Act named this site, most recently on May 14, 2026.

  • Inspection ending Jun 13, 2025VAI

    3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 5, 2025. FDAgent holds no copy that FDA posted.

    1. May 14, 2026FDA-FOIA-2026-4951Form 483· 1 of 5 sitesRecords released
    2. Nov 17, 2025FDA-FOIA-2025-13042Form 483· 1 of 2 sitesRecords released
    3. Aug 5, 2025FDA-FOIA-2025-9467Form 483· 1 of 4 sitesRecords released
  • Inspection ending Mar 28, 2019NAI

    1 request. In the inspection history above.

    1. Aug 9, 2019FDA-FOIA-2019-7194EIROpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 7 inspections, with how many of them cited each.

Show all 11

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1718912, one row per refused product line.

9 refusals from 2021-02-19 to 2026-03-12 · charged FRNMFGREG on 7, NOT LISTED on 6 · no FDA sample analysis

FDA firm profile
  • 2026-03-12

    DENTAL PROTECTANT N.E.C.63SBL99

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southwest ImportsSCS-9225990-8/20001/1

  • 2026-03-12

    DENTAL PROTECTANT N.E.C.63SBL99

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southwest ImportsSCS-9225990-8/10001/1

  • 2022-04-14

    CORD, RETRACTION76MVL

    RXPERSONAL 502(a), 502(f)(1), 801(a)(3), MISBRANDING

    Division of Southeast Imports799-8390068-8/11/1

  • 2021-08-30

    SEALANT, PIT AND FISSURE, AND CONDITIONER76EBC

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsSCS-7217784-5/11/1

  • 2021-06-14

    ORTHODONTIC LED ACCESSORY76PLH

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-2306166-6/11/1

Show 4 more
  • 2021-02-19

    STRIP, POLISHING AGENT76EHM

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsSCS-3562489-1/491/1

  • 2021-02-19

    RESTORATION, RESIN76NSM

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsSCS-3562489-1/481/1

  • 2021-02-19

    STRIP, POLISHING AGENT76EHM

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsSCS-3562489-1/201/1

  • 2021-02-19

    STRIP, POLISHING AGENT76EHM

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsSCS-3562489-1/191/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1718912 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1718912. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.