Skip to content

Etymotic Research, Inc.

1125 Remington Rd, Schaumburg, Illinois, United States

FEI 1450042

FDA has 4 inspections on file for Etymotic Research, Inc. in Schaumburg, Illinois (FEI 1450042), from 2009 to 2017. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2009–2017
0 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
2 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Management responsibility

      Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.

    2. Changes to correct/prevent quality problems

      Changes in methods and procedures needed to correct and prevent identified quality problems are not implemented.

    3. Identification of actions needed

      Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.

    4. Uniform and timely processing

      Complaint handling procedures have not been defined to ensure that all complaints are processed in a uniform and timely manner.

    5. Procedure for analysis of data sources

      The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not defined and implemented.

    Show 6 more citations
    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance or rejection of incoming product were not established, defined, documented, complete, and implemented.

    2. ID label, labeling

      The device history record does not include the primary identification label and labeling for each device.

    3. Statistical methodology

      Corrective and preventive action procedures addressing the use of appropriate statistical methodology to identify existing and potential causes of nonconforming product or other quality problems were not established, defined, and complete.

    4. Lack of or inadequate In-process acceptance procedures

      Acceptance procedures to ensure that in-process product is controlled until the required inspection and tests have been completed and approved were not documented, complete, and implemented.

    5. Verification vs validation

      There is no corrective and preventive action procedure to determine when verification can be conducted in lieu of validation.

    6. DHR content - manufacturing dates

      The device history record does not include manufacturing dates.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not established, defined, documented, complete, and implemented.

    2. Software validation for automated processes

      Software used as part of the quality system has not been validated for its intended use according to an established protocol.

    3. Quality Audit/Reaudit - conducted

      Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.

    4. Specific non-conforming product procedures

      Procedures for addressing the identification, documentation, evaluation, segregation, disposition, and investigation of nonconforming product were not complete.

    5. Not approved or obsolete document retrieval

      Documents that were not approved were observed at a location where they are being used.

    Show 5 more citations
    1. Lack of or inadequate DHR procedures

      Procedures have not been implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.

    2. General

      Complaint handling procedures for complaints have not been completed and implemented.

    3. Management review - Lack of or inadequate procedures

      Procedures for management review are not established.

    4. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not complete and implemented.

    5. Lack of procedures, or not maintained

      Document control procedures were not complete.

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.

Show all 9

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1450042 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1450042. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.