Skip to content

Medela Inc

1101 Corporate Dr, Mchenry, Illinois, United States

FEI 1419937

FDA has 8 inspections on file for Medela Inc in Mchenry, Illinois (FEI 1419937), from 2009 to 2014. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on May 25, 2012. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.30(d) and 21 CFR 820.30(g).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
8, 2009–2014
2 OAI · 3 VAI · 3 NAI
Inspections with 483 observations shown
3 of 8
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Design validation - production units

      The design was not validated.

    5. Design output - documentation

      Design output was not adequately documented before release.

    Show 1 more citation
    1. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification - documentation

      The design verification results, including identification of the design, methods, the date, and the individual performing the verification, were not documented and filed in the design history file.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures were not followed for the identification, documentation, validation or verification, review, and approval of design changes before their implementation.

    4. Identification of actions needed

      Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.

    5. Analysis of data/reports from data sources

      Not all data from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.

    Show 2 more citations
    1. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    2. Uniform and timely processing

      Complaint handling procedures have not been defined and implemented to ensure that all complaints are processed in a uniform and timely manner.

  8. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities have not been documented, including dissemination of information about quality problems or nonconforming product to responsible parties.

    2. Design validation - documentation

      The results of design validation, including identification of the design, methods, the date, and the individual performing the validation, were not documented and filed in the design history file.

    3. Essential design outputs

      Design outputs that are essential for the proper functioning of the device were not completely identified.

    4. Design control - no procedures

      Procedures to control the design process of the device were not established, defined, documented, and complete.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 8 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1419937 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1419937. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.