FDA has 3 inspections on file for Oak Ridge Products, LLC in Mchenry, Illinois (FEI 1419823), from 2017 to 2026. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA posted the Form 483 from that inspection. FDA published 6 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.198(a).

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2017–2026
0 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
↗
Requested under FOIA
3 requests, latest Jul 7, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesPosted 483 · 6 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.3.9

      Records of design and development changes and their reviews not maintained

      Records of design and development changes, the review of design and development changes and necessary actions associated with design and development changes were not adequately maintained.

    2. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures not documented

      Procedures for timely complaint handling have not been adequately documented.

    3. ISO 13485:2016 Clause 8.2.4

      Internal audits not conducted to ensure QMS meets requirements and is effective

      Internal audits have not been adequately conducted at planned intervals to determine whether the QMS conforms to planned and documented arrangements and applicable requirements and is effectively implemented and maintained.

    4. ISO 13485:2016 Clause 5.6.1

      QMS not reviewed by top management at documented planned intervals

      Top management did not adequately review the quality management system at documented planned intervals to ensure its continued suitability, adequacy, and effectiveness.

    5. ISO 13485:2016 Clause 8.3.1

      Procedure for nonconforming product not documented

      Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not adequately documented.

    Show 1 more citation
    1. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been adequately documented.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    2. Design control - no procedures

      Procedures for design control have not been established.

    3. Management review dates

      The results and/or dates of management reviews are not documented.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 3 more citations
    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    2. Training records

      Personnel training is not documented.

    3. Quality Audits - defined intervals

      Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Jul 7, 2026. Each one asked for a Form 483.

  • Inspection ending Apr 1, 2026NAI

    3 requests. In the inspection history above. FDA has posted its Form 483.

    1. Jul 7, 2026FDA-FOIA-2026-6458Form 483· 1 of 4 sitesRecords released
    2. Jun 19, 2026FDA-FOIA-2026-5955Form 483· 1 of 5 sitesRecords released
    3. Apr 28, 2026FDA-FOIA-2026-4216Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1419823 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1419823. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.