Skip to content

Oak Ridge Products, LLC Form 483, April 2026

FDA issued this Form 483 to Oak Ridge Products, LLC in Mchenry, Illinois (FEI 1419823) at the end of a device inspection that ended April 1, 2026. FDA classified the inspection No Action Indicated (NAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 13, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Apr 1, 2026
Devices · Postmarket Assurance: Devices
FDA's outcome
NAI
No Action Indicated
Posted by FDA
Nov 13, 2026
226 days after the inspection
Facility
FEI 1419823
Mchenry, Illinois

FDA's 6 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. ISO 13485:2016 Clause 7.3.9

    Records of design and development changes and their reviews not maintained

    Records of design and development changes, the review of design and development changes and necessary actions associated with design and development changes were not adequately maintained.

  2. ISO 13485:2016 Clause 8.2.2

    Complaint handling procedures not documented

    Procedures for timely complaint handling have not been adequately documented.

  3. ISO 13485:2016 Clause 8.2.4

    Internal audits not conducted to ensure QMS meets requirements and is effective

    Internal audits have not been adequately conducted at planned intervals to determine whether the QMS conforms to planned and documented arrangements and applicable requirements and is effectively implemented and maintained.

  4. ISO 13485:2016 Clause 5.6.1

    QMS not reviewed by top management at documented planned intervals

    Top management did not adequately review the quality management system at documented planned intervals to ensure its continued suitability, adequacy, and effectiveness.

  5. ISO 13485:2016 Clause 8.3.1

    Procedure for nonconforming product not documented

    Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not adequately documented.

  6. ISO 13485:2016 Clause 8.5.2

    Corrective action procedures not documented

    Corrective action procedures defining the requirements for verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been adequately documented.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.