Tetra Medical Supply Corp.
6364 W Gross Point Rd, Niles, Illinois, United States
FEI 1418270
FDA has 2 inspections on file for Tetra Medical Supply Corp. in Niles, Illinois (FEI 1418270), from 2016 to 2017. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Apr 22, 2016. No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.75(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2016–2017
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier oversight
The type and extent of control to be exercised over suppliers was not clearly defined.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documented evaluation
The evaluation of potential suppliers and contractors was not documented.
Lack of Written MDR Procedures
Written MDR procedures have not been developed and implemented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 4 more citations
Quality plan
A quality plan has not been established.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a method and process have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been established.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 820.75Process validation2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.184Device history record1 of 2
Show all 9
- 21 CFR 820.198Complaint files1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1418270 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1418270. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.