Skip to content

Vilex in Tennessee, Inc.

111 Moffitt St, Mcminnville, Tennessee, United States

FEI 1051526

FDA has 4 inspections on file for Vilex in Tennessee, Inc. in Mcminnville, Tennessee (FEI 1051526), from 2010 to 2018. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.30(g) and 21 CFR 820.30(b).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2010–2018
0 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been documented in the device history record.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Design validation - Risk analysis

      Results of the design risk analysis were not adequately documented.

    4. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    5. Design history file

      The design history file was not established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    2. Design input - documentation

      Design input requirements were not adequately documented.

    3. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

    4. Design plans - Lack of or inadequate

      Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    Show 2 more citations
    1. Design review - Lack of or inadequate procedures

      Procedures for design review have not been adequately established.

    2. Design transfer - Lack of or inadequate procedures

      Procedures for design transfer have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation, timeliness, identification

      Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.

    2. Design plans - Lack of or inadequate

      Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1051526 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1051526. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.