Amerx Health Care Corp.
1300 S Highland Ave, Clearwater, Florida, United States
FEI 1000525408
FDA has 2 inspections on file for Amerx Health Care Corp. in Clearwater, Florida (FEI 1000525408), from 2014 to 2016. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 8 citations from this inspection: 7 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
The most recent OAI inspection ended on Feb 24, 2014. No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2014–2016
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier oversight
The type and extent of control to be exercised over the product , services , suppliers , contractors and consultants was not clearly defined.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Complaints
Complaint files are not maintained. Specifically ***
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification , communication and evaluation of events that may be subject to medical device reporting requirements.
Show 3 more citations
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Approval, inadequate purchasing data
Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been approved , adequately approved and adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices19 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Design output - Lack of or inadequate procedures
Procedures for design output have not been established.
Complaints
Complaint files are not maintained.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Show 14 more citations
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been established.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.
Design history file
The design history file was not established.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
Design review - Lack of or inadequate procedures
Procedures for design review have not been established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained, and implemented.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Design input - Lack of or inadequate procedures
Procedures for design input have not been established.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 820.90Nonconforming product2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
Show all 9
- 21 CFR 820.184Device history record1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000525408 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000525408. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.