Skip to content

Mirac LLC

2153 Winners Cir, Dayton, Ohio, United States

FEI 1000160007

FDA has 2 inspections on file for Mirac LLC in Dayton, Ohio (FEI 1000160007), from 2023 to 2026. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA posted the Form 483 from that inspection. FDA published 9 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2023–2026
1 OAI · 1 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
↗
Requested under FOIA
2 requests, latest Jun 18, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesPosted 483 · 9 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 8.1

      Monitoring, measurement, analysis, and improvement processes not adequate

      The monitoring, measurement, analysis, and improvement processes were not adequately planned and implemented to demonstrate conformity of product, ensure conformity of the quality management system and maintain the effectiveness of the quality management system.

    2. ISO 13485:2016 Clause 4.2.2

      Specific requirements of the Quality Manual not documented

      The quality manual does not adequately include description of the interaction between processes of the quality management system.

    3. ISO 13485:2016 Clause 8.3.1

      Procedure for nonconforming product not documented

      Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not documented.

    4. ISO 13485:2016 Clause 7.2.2

      Required actions not taken when product requirements changed

      When product requirements were changed, your firm did not ensure relevant documents were adequately amended and relevant personnel were made aware of the changes.

    5. ISO 13485:2016 Clause 7.4.1

      Records of supplier (re)evaluation, selection, and monitoring not maintained

      Records of the results of evaluation, selection, monitoring and re-evaluation of supplier capability or performance were not adequately maintained.

    Show 4 more citations
    1. ISO 13485:2016 Clause 4.1.2

      Risk-based approach not applied to QMS processes

      A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.

    2. ISO 13485:2016 Clause 5.5.1

      Management responsibilities and authorities not documented

      Top management did not ensure that responsibilities and authorities are adequately defined, documented, and communicated within the organization.

    3. ISO 13485:2016 Clause 7.5.6

      Process for production and service provision not validated

      Processes for production and service provision where the resulting output cannot be, or is not, verified by subsequent monitoring or measurement were not adequately validated.

    4. ISO 13485:2016 Clause 8.2.4

      Internal audit procedures not documented

      Procedures to describe the responsibilities and requirements for planning and conducting internal audits and recording and reporting audit results were not adequately documented.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Jun 18, 2026. Each one asked for a Form 483.

  • Inspection ending Feb 17, 2026OAI

    2 requests. In the inspection history above. FDA has posted its Form 483.

    1. Jun 18, 2026FDA-FOIA-2026-5913Form 483Records released
    2. Apr 23, 2026FDA-FOIA-2026-4082Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 5 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000160007 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000160007. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.