Mirac LLC Form 483, February 2026
FDA issued this Form 483 to Mirac LLC in Dayton, Ohio (FEI 1000160007) at the end of a device inspection that ended February 17, 2026. FDA classified the inspection Official Action Indicated (OAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Nov 13, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Feb 17, 2026
- Devices · Compliance: Devices
- FDA's outcome
- OAI
- Official Action Indicated
- Posted by FDA
- Nov 13, 2026
- 269 days after the inspection
FDA's 9 citations from this inspection
FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.
- ISO 13485:2016 Clause 8.1
Monitoring, measurement, analysis, and improvement processes not adequate
The monitoring, measurement, analysis, and improvement processes were not adequately planned and implemented to demonstrate conformity of product, ensure conformity of the quality management system and maintain the effectiveness of the quality management system.
- ISO 13485:2016 Clause 4.2.2
Specific requirements of the Quality Manual not documented
The quality manual does not adequately include description of the interaction between processes of the quality management system.
- ISO 13485:2016 Clause 8.3.1
Procedure for nonconforming product not documented
Procedures to define the controls and related responsibilities and authorities for the identification, documentation, segregation, evaluation, and disposition of nonconforming product were not documented.
- ISO 13485:2016 Clause 7.2.2
Required actions not taken when product requirements changed
When product requirements were changed, your firm did not ensure relevant documents were adequately amended and relevant personnel were made aware of the changes.
- ISO 13485:2016 Clause 7.4.1
Records of supplier (re)evaluation, selection, and monitoring not maintained
Records of the results of evaluation, selection, monitoring and re-evaluation of supplier capability or performance were not adequately maintained.
- ISO 13485:2016 Clause 4.1.2
Risk-based approach not applied to QMS processes
A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.
- ISO 13485:2016 Clause 5.5.1
Management responsibilities and authorities not documented
Top management did not ensure that responsibilities and authorities are adequately defined, documented, and communicated within the organization.
- ISO 13485:2016 Clause 7.5.6
Process for production and service provision not validated
Processes for production and service provision where the resulting output cannot be, or is not, verified by subsequent monitoring or measurement were not adequately validated.
- ISO 13485:2016 Clause 8.2.4
Internal audit procedures not documented
Procedures to describe the responsibilities and requirements for planning and conducting internal audits and recording and reporting audit results were not adequately documented.
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.