LeMaitre Vascular, Inc.
206 N Center Dr, North Brunswick, New Jersey, United States
FEI 1000121696
FDA has 8 inspections on file for LeMaitre Vascular, Inc. in North Brunswick, New Jersey (FEI 1000121696), from 2010 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 8 citations from this inspection. They are listed below. Two of the four FOIA requests for this site have asked FDA for this inspection's 483.
The most recent OAI inspection ended on Mar 8, 2010. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
Eight regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 8, 2010–2026
- 1 OAI · 2 VAI · 5 NAI
- Inspections with 483 observations shown
- 6 of 8
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Aug 21, 2026
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.5.11
Procedures for preserving the conformity of product not documented
Procedures were not adequately documented for preserving the conformity of product to requirements during processing, storage, handling, and distribution.
- ISO 13485:2016 Clause 7.3.3
Design and development inputs do not include all requirements
Design and development inputs related to product requirements do not adequately include requirements essential for design and development of the product and processes.
- ISO 13485:2016 Clause 8.2.1
Information gathered in feedback process not used as input into risk management
Information gathered in the feedback process does not serve as potential input into risk management for monitoring and maintaining product requirements as well as the product realization or improvement processes.
- ISO 13485:2016 Clause 6.4.2
Arrangements for control of contaminated product not planned and documented
Arrangements were not adequately planned and documented for the control of contaminated or potentially contaminated product in order to prevent contamination of the work environment, personnel, or product.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device have not been documented.
Show 3 more citations
- ISO 13485:2016 Clause 8.5.3
Preventive action procedures not documented
Preventive action procedures describing the requirements for planning and documenting action needed and implementing such action have not been adequately documented.
- ISO 13485:2016 Clause 7.5.1
Production and service provision not adequate
Production and service provision was not adequately planned, carried out, monitored, and controlled to ensure that product conforms to specification.
- ISO 13485:2016 Clause 8.3.1
Nonconforming product not adequately evaluated
Nonconforming product was not evaluated to determine the need for an investigation and notification of any external party responsible for the nonconformity.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Show 3 more citations
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Sampling plans
Sampling plans are not based on valid statistical rationale.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Documentation - review in response to changes or deviations
There is no documentation of the revalidation of a process performed in response to changes or process deviations.
Documentation
Process validation have not been adequately documented and adequately approved.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Environmental control Lack of or inadequate procedures
Appropriate procedures have not been defined and followed for controlling environmental conditions.
Procedure for analysis of data sources
The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not complete.
Storage procedures to avoid release of unsuitable product
Procedures for controlling the storage of product in storage areas and stock rooms are not adequate to ensure that no obsolete, rejected, or deteriorated product is distributed.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately and fully validated and approved according to established procedures.
Documentation - review in response to changes or deviations
There is no documentation of the revalidation of a process conducted in response to changes or process deviations.
Show 3 more citations
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Contamination control, Lack of or inadequate procedures
Procedures have not been defined and followed to prevent contamination of equipment or product by certain substances.
Nonconforming product evaluation/investigation
The evaluation of nonconforming product has not been documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Aug 21, 2026. Each one asked for a Form 483.
Inspection ending Jun 25, 2026NAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 3, 2026. FDAgent holds no copy that FDA posted.
- Aug 21, 2026FDA-FOIA-2026-8140Form 483· 1 of 2 sitesRecords released
- Aug 12, 2026FDA-FOIA-2026-7776Form 483· 1 of 5 sitesRecords released
Inspection ending Apr 16, 2025NAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Oct 6, 2025. FDAgent holds no copy that FDA posted.
- Aug 26, 2025FDA-FOIA-2025-10191Form 483Records released
- Jun 25, 2025FDA-FOIA-2025-8474Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.75Process validation5 of 6
- 21 CFR 820.100Corrective and preventive action4 of 6
- 21 CFR 820.70Production and process controls3 of 6
- 21 CFR 820.25Personnel2 of 6
- 21 CFR 820.184Device history record2 of 6
- 21 CFR 820.30Design controls1 of 6
- 21 CFR 820.90Nonconforming product1 of 6
- 21 CFR 820.150Storage1 of 6
Show all 10
- 21 CFR 820.181Device master record1 of 6
- 21 CFR 820.250Statistical techniques1 of 6
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Other FEIs recorded as LeMaitre Vascular, Inc.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
LeMaitre Vascular, Inc.FEI 1220948Burlington, Massachusetts, United States
Last inspected 2023-05-16VAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000121696 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000121696. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.