Skip to content

Micro Measurement Laboratories, Inc.

1300 S Wolf Rd, Wheeling, Illinois, United States

FEI 1000115216DUNS 119461311

FDA has 9 inspections on file for Micro Measurement Laboratories, Inc. in Wheeling, Illinois (FEI 1000115216), from 2009 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 211.194(a)(2).

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for analysis, through Dec 31, 2027. Its registrant is Gillson Sciences, Inc. (D-U-N-S 119461256).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
9, 2009–2026
0 OAI · 5 VAI · 4 NAI
Inspections with 483 observations shown
4 of 9
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Jun 5, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

9 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Computer control of master formula records

      Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel.

    2. SOPs not followed / documented

      Written production and process control procedures are not followed in the execution of production and process control functions.

    3. GMP Training Frequency

      GMP training is not conducted to assure that employees remain familiar with CGMP requirements applicable to them.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  3. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Acceptance of drug products

      Determinations of conformance to appropriate written specifications for acceptance are deficient for drug products.

    2. Suitability of testing methods verified

      The suitability of all testing methods is not verified under actual conditions of use.

    3. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Suitability of testing methods verified

      The suitability of all testing methods is not verified under actual conditions of use.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    2. Written record of investigation incomplete

      Written records of investigations into unexplained discrepancies and the failure of a batch or any of its components to meet specifications do not always include the conclusions and follow-up.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  6. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  7. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  8. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  9. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Jun 5, 2026. It asked for a Form 483.

  • Inspection ending Apr 28, 2026VAI

    1 request. In the inspection history above.

    1. Jun 5, 2026FDA-FOIA-2026-5585Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 9 inspections, with how many of them cited each.

Show all 9

FDA drug establishment registration

FDA registration search
Registered operations
Analysis
Registration runs through
Dec 31, 2027
Registrant
Gillson Sciences, Inc.D-U-N-S 119461256

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000115216 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000115216. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.