Skip to content

CareFusion 303, Inc. Form 483, April 2020

FDA issued this Form 483 to CareFusion 303, Inc. in San Diego, California (FEI 2016493) at the end of a device inspection that ended April 6, 2020. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 23, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Apr 6, 2020
Devices · Compliance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Oct 23, 2026
2,391 days after the inspection
Facility
FEI 2016493
San Diego, California

FDA's 8 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Investigation of device failures

    Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

  2. Lack of Written MDR Procedures

    Written MDR procedures have not been implemented.

  3. Lack of or inadequate procedures

    Procedures for corrective and preventive action have not been adequately established.

  4. Design validation - software validation not performed

    Validation of device software is incomplete.

  5. Design verification - Lack of or inadequate procedures

    Procedures for design verification have not been adequately established.

  6. Servicing - Lack of or inadequate procedures

    Procedures or instructions for performing servicing activities have not been adequately established.

  7. Lack of or inadequate complaint procedures

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  8. Report of violation of the Act (see 803.52(e)(9))

    A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.

Other Form 483s FDA posted for this facility

Every Form 483 FDA posts under FEI 2016493, newest first.

All inspections at this facility →

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.