Skip to content

Hudson Scientific LLC

103 W Main St, Hudson, Michigan, United States

FEI 3013037032DUNS 080339637

FDA has 2 inspections on file for Hudson Scientific LLC in Hudson, Michigan (FEI 3013037032), from 2018 to 2020. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, labeling, relabeling and repacking, through Dec 31, 2026. Its registrant is Hudson Scientific LLC (D-U-N-S 080339637).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2018–2020
0 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    2. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    3. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Calibration/Inspection/Checking not done

      Routine checking of electronic equipment is not performed according to a written program designed to assure proper performance.

    2. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not fully followed.

    3. Written sanitation procedures not followed

      Written procedures for sanitation are not followed.

    4. Lack of written stability program

      There is no written testing program designed to assess the stability characteristics of drug products.

    5. SOPs not followed / documented

      Written production and process control procedures are not documented at the time of performance.

    Show 1 more citation
    1. Procedures To Be in Writing

      Written procedures are lacking which describe in sufficient detail the receipt, identification, storage, handling, sampling, testing, approval and rejection of components, drug product containers and closures.

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Show all 9

FDA drug establishment registration

FDA registration search
Registered operations
Manufacturing, labeling, relabeling, repacking
Registration runs through
Dec 31, 2026
Registrant
Hudson Scientific LLCD-U-N-S 080339637

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3013037032 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3013037032. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.