Robin Industries, Inc.
3500 State Route 39, Millersburg, Ohio, United States
FEI 3011672474
FDA has 4 inspections on file for Robin Industries, Inc. in Millersburg, Ohio (FEI 3011672474), from 2016 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection. They are listed below. Two of the four FOIA requests for this site have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.72(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2016–2026
- 0 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 4 requests, latest Oct 16, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 6.3
Requirements for infrastructure maintenance activities not documented
Requirements for infrastructure maintenance activities that can affect product quality, including the interval of performing the maintenance activities, have not been documented.
- ISO 13485:2016 Clause 6.4.1
Work environment requirements and procedures not documented
Requirements for the work environment, and the procedures to monitor and control the work environment, were not adequately documented where conditions for the work environment can have an adverse effect on product quality.
- ISO 13485:2016 Clause 8.1
Methods for monitoring, measurement, analysis, and improvement not determined
Appropriate methods, including statistical techniques, and the extent of their use were not adequately determined for monitoring, measurement, analysis, and improvement processes.
- ISO 13485:2016 Clause 7.1
Processes needed for product realization not planned and developed
Processes needed for product realization were not adequately planned and developed.
- ISO 13485:2016 Clause 4.2.4
Lack of document control
Documents required by the quality management system have not been adequately controlled.
Show 1 more citation
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for reviewing nonconformities, determining the causes of nonconformities, evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action and verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device have not been adequately documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Equipment control activity documentation
Equipment maintenance activities have not been documented.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they could be used.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Acceptable supplier records, inadequate records
Records of acceptable suppliers have not been adequately established.
Production processes
Production processes were not controlled to ensure that a device conforms to its specifications.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Four requests to FDA under the Freedom of Information Act named this site, most recently on Oct 16, 2026. Each one asked for a Form 483.
Inspection ending Jul 24, 2026VAI
2 requests. In the inspection history above.
- Oct 16, 2026FDA-FOIA-2026-10084Form 483· 1 of 5 sitesOpen
- Sep 29, 2026FDA-FOIA-2026-9475Form 483· 1 of 4 sitesOpen
Inspection ending Sep 26, 2023VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jan 2, 2024FDA-FOIA-2024-17Form 483Closed
- Oct 31, 2023FDA-FOIA-2023-9547Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 3
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
- 21 CFR 820.184Device history record1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3011672474 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3011672474. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.