Forward Science Technologies LLC
10810 Craighead Dr, Houston, Texas, United States
FEI 3010143501DUNS 079212068
FDA has 5 inspections on file for Forward Science Technologies LLC in Houston, Texas (FEI 3010143501), from 2014 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the three FOIA requests for this site has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing and repacking, through Dec 31, 2027. Its registrant is Forward Science Technologies LLC (D-U-N-S 079212068).
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2014–2025
- 0 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest May 15, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Design validation acceptance criteria
Acceptance criteria were not established prior to the performance of validation activities.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Show 1 more citation
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complaint Handling Procedure
Procedures describing the handling of written and oral complaints related to drug products are not written or followed.
Written procedures not established/followed
Written procedures are not established for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product.
Handling and Storage to Prevent Contamination
There was a failure to handle and store components, drug product containers and closures at all times in a manner to prevent contamination.
Sanitation--buildings not clean, free of infestation
Buildings used in the manufacture, processing, packing or holding of drug products are not maintained in a clean and sanitary condition.
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Show 1 more citation
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on May 15, 2026. Each one asked for a Form 483.
Inspection ending Jun 12, 2025VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 12, 2025. FDAgent holds no copy that FDA posted.
- Aug 5, 2025FDA-FOIA-2025-9490Form 483· 1 of 5 sitesRecords released
Inspection ending Mar 26, 2025VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on May 21, 2026. FDAgent holds no copy that FDA posted.
- May 15, 2026FDA-FOIA-2026-5000Form 483· 1 of 5 sitesRecords released
- May 22, 2025FDA-FOIA-2025-4543Form 483· 1 of 5 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 211.56Sanitation1 of 3
- 21 CFR 211.67Equipment cleaning and maintenance1 of 3
- 21 CFR 211.80General requirements1 of 3
- 21 CFR 211.198Complaint files1 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.75Process validation1 of 3
Show all 10
- 21 CFR 820.90Nonconforming product1 of 3
- 21 CFR 820.198Complaint files1 of 3
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, repacking
- Registration runs through
- Dec 31, 2027
- Registrant
- Forward Science Technologies LLCD-U-N-S 079212068
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3010143501 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3010143501. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.