3m Canada Company
60 California Avenue, Brockville, Canada
FEI 3009699611
FDA has one inspection on file for 3m Canada Company in Brockville, Canada (FEI 3009699611). It ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 1, in 2015
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a process have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.70Production and process controls1 of 1
- 21 CFR 820.90Nonconforming product1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.198Complaint files1 of 1
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3009699611, one row per refused product line.
1 refusal on 2023-10-13 · charged NOT LISTED on 1, REGISTERED on 1 · no FDA sample analysis
FDA firm profile- 2023-10-13
BARRIER FACE COVERING PER ENFORCEMENT POLICY FOR FACE MASKS AND RESPIRATORS DURING THE COVID-19 PUBL80QOZ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Northern Border Imports9RK-4020892-8/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3009699611 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3009699611. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.