Skip to content

Ocusun, LLC

150 Alameda Del Prado, Novato, California, United States

FEI 3007289540

FDA has 3 inspections on file for Ocusun, LLC in Novato, California (FEI 3007289540), from 2009 to 2015. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.22.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2009–2015
0 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
2 of 3
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not defined.

    2. Review and evaluation for investigation

      Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.

    3. Training - Lack of or inadequate procedures

      Procedures for identifying training needs were not established.

    4. Investigation

      Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.

    5. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not defined.

    Show 2 more citations
    1. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by suppliers and consultants were not .

    2. Quality Audit/Reaudit - conducted

      Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality plan

      A quality plan was not defined, documented, and implemented.

    2. General

      Complaint handling procedures for reviewing complaints have not been established, defined, and documented.

    3. Control/distribution procedures

      Procedures for the control and distribution of finished devices have not been defined, documented, and implemented to ensure that only devices approved for release are distributed.

    4. Management review - Lack of or inadequate procedures

      Procedures for management review are not defined, documented, and implemented.

    5. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not defined and documented.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007289540 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007289540. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.