Ocular Systems, Inc.
15201 E Moncrieff Pl Ste C, Aurora, Colorado, United States
FEI 3006480421
FDA has 2 inspections on file for Ocular Systems, Inc. in Aurora, Colorado (FEI 3006480421), from 2009 to 2010. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 10 citations from this inspection, all under 21 CFR Part 1271, the rules for human cell and tissue products. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2009–2010
- 0 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Validation for intended use
The performance of commercial computer software that has been customized or programmed, including changes to computer software, has not been validated for the intended use, when such software is relied upon to comply with core GTP requirements.
Qualifications lacking
Personnel do not have the necessary training to ensure competent performance of their assigned functions.
Name and address on summary
The summary of records for HCT/Ps did not contain the address of the establishment that made the donor-eligibility determination.
Adequate control, proper conditions
Environmental conditions existed in which contamination or cross contamination of HCT/Ps or equipment could occur, and environmental conditions were not adequately controlled and proper conditions for operations were not provided.
Shipping conditions appropriate
Appropriate shipping conditions were not established for each type of HCT/P.
Show 5 more citations
Aseptic processing---cleaning, disinfecting
Environmental controls do not provide for adequate cleaning and disinfecting of rooms and equipment to ensure aseptic processing.
Procedures and release criteria
Procedures including release criteria for activities relating to the shipment and distribution of HCT/Ps were not established.
Verification/validation
Process changes were not verified to ensure the change does not cause an adverse impact elsewhere in operations.
Monitoring systems
The quality program has not established and maintained appropriate monitoring systems.
Validation & approval--established procedures
Processes with results which could not be fully verified by inspection and tests, were not validated and approved according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies23 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Acceptance criteria designed to prevent CD
Incoming HCT/Ps were not accepted, rejected, and placed in quarantine based on pre-established criteria designed to prevent communicable disease transmission.
Verification/validation
Process changes were not validated to ensure the change does not cause an adverse impact elsewhere in operations.
Process validation procedures
Procedures to validate and approve processes that cannot be fully verified by inspection and tests were not established, implemented, reviewed, and revised.
Deviations not reported to FDA
HCT/P deviations relating to core CGTP requirements that occurred in your establishment were not reported to FDA.
Adequate control, proper conditions
Environmental conditions existed in which contamination or cross contamination of HCT/Ps or equipment could occur, and environmental conditions were not adequately controlled.
Show 18 more citations
Records maintained concurrently
Records were not maintained concurrently with the performance of each step.
Calibration procedures and schedules (general)
The schedules for the calibration of equipment used for measuring were not followed.
Procedures re complaints
Procedures for the documentation and investigation of complaints relating to CGTP requirements were not established and maintained.
Monitoring--cross contamination, exposure to CD
Environmental conditions are not monitored when such conditions could cause contamination or cross contamination of HCT/Ps or equipment.
Causing contamination, increasing risks
HCT/Ps were not processed in a way that does not cause contamination or cross contamination during processing and that does not increase the risk of introduction, transmission, or spread of communicable disease.
Contamination, mix ups, improper release
Storage areas and stock rooms were not controlled to prevent mix-ups, contamination and cross contamination of HCT/Ps, supplies and reagents.
Controlling the labeling of HCT/Ps
Procedures to control the labeling of HCT/Ps were not established, maintained, defined, and followed.
Summary--records used to make determination
After completion of the donor-eligibility determination, HCT/Ps were not accompanied with the summary of the records used to make the donor-eligibility determination.
All labeling requirements met
Labeling procedures did not ensure that HCT/Ps are labeled in accordance with all labeling requirements.
Use prior to verification
Supplies and reagents were used before they were verified to meet specifications designed to prevent the introduction, transmission, or spread of communicable disease.
Deviations--Investigation, documenting, trending
The quality program does not include the investigation, the trending, and the reporting of HCT/P deviations relating to core CTGP requirements.
Validation for intended use
The performance of commercial computer software that has been customized or programmed, including changes to computer software, has not been validated for the intended use, when such software is relied upon to comply with core GTP requirements.
Computer software verification
The performance of computer software has not been verified.
Aseptic processing---cleaning, disinfecting
Environmental controls do not provide for adequate cleaning and disinfecting of rooms to ensure aseptic processing.
Separate or defined operational areas
Operations were not performed in a facility with separate or defined areas of adequate size to prevent contamination, cross contamination, and accidental exposure of HCT/Ps to communicable disease agents.
Design appropriate, suitably located
Equipment used in the manufacture of HCT/Ps was not of appropriate design for its use.
Contract firms--identities and responsibilities
The list of responsibilities of establishment(s) that perform manufacturing steps for you was not maintained.
Procedures and release criteria
Procedures including release criteria for activities relating to the shipment and distribution of HCT/Ps were not established and maintained.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 1271.55What records must accompany an HCT/P after the donor-eligibility determination is complete; and what records must I retain?2 of 2
- 21 CFR 1271.160Establishment and maintenance of a quality program2 of 2
- 21 CFR 1271.195Environmental control and monitoring2 of 2
- 21 CFR 1271.225Process changes2 of 2
- 21 CFR 1271.230Process validation2 of 2
- 21 CFR 1271.265Receipt, predistribution shipment, and distribution of an HCT/P2 of 2
- 21 CFR 1271.170Personnel1 of 2
- 21 CFR 1271.190Facilities1 of 2
Show all 16
- 21 CFR 1271.200Equipment1 of 2
- 21 CFR 1271.210Supplies and reagents1 of 2
- 21 CFR 1271.220Processing and process controls1 of 2
- 21 CFR 1271.250Labeling controls1 of 2
- 21 CFR 1271.260Storage1 of 2
- 21 CFR 1271.270Records1 of 2
- 21 CFR 1271.320Complaint file1 of 2
- 21 CFR 1271.350Reporting1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006480421 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006480421. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.