Nutronix Revolution, Inc.
7308 Hanover Green Dr Ste 300, Mechanicsville, Virginia, United States
FEI 3005984610
FDA has one inspection on file for Nutronix Revolution, Inc. in Mechanicsville, Virginia (FEI 3005984610). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 16 citations from this inspection: 10 under 21 CFR Part 211, the manufacturing rules for finished drugs, 5 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2015
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained, and implemented.
Training--operations, GMPs, written procedures
Employees are not given training in the particular operations they perform as part of their function and current good manufacturing practices.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing.
Show 11 more citations
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
Prepared for each batch, include complete information
Batch production and control records are not prepared for each batch of drug product produced.
Lack of written stability program
There is no written testing program designed to assess the stability characteristics of drug products.
Test methods
The accuracy, sensitivity, specificity, and reproducibility of test methods have not been established and documented.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Written procedures describing in detail
There is a lack of written procedures describing in sufficient detail the receipt, identification, storage, handling, sampling, examination, and testing of labeling and packaging materials.
Absence of Written Procedures
There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess.
Recall facilitation
A system by which the distribution of each lot of drug product can be readily determined to facilitate its recall if necessary, has not been established.
Complaint Handling Procedure
Procedures describing the handling of written and oral complaints related to drug products are not written or followed.
Supervisor Training/Education/Experience
Individuals responsible for supervising the manufacture, processing, packing, and holding of a drug product lack the training and experience to perform their assigned functions in such a manner as to assure the drug product has the safety, identity, strength, quality and purity that it purports or is represented to possess.
Quality System Procedures
Quality system procedures and instructions have not been established.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 211.22Responsibilities of quality control unit1 of 1
- 21 CFR 211.25Personnel qualifications1 of 1
- 21 CFR 211.100Written procedures; deviations1 of 1
- 21 CFR 211.122Materials examination and usage criteria1 of 1
- 21 CFR 211.150Distribution procedures1 of 1
- 21 CFR 211.165Testing and release for distribution1 of 1
- 21 CFR 211.166Stability testing1 of 1
- 21 CFR 211.188Batch production and control records1 of 1
Show all 15
- 21 CFR 211.198Complaint files1 of 1
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.70Production and process controls1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.198Complaint files1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005984610 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005984610. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.