Brava, LLC
14221 SW 142nd St, Miami, Florida, United States
FEI 3005643748
FDA has 4 inspections on file for Brava, LLC in Miami, Florida (FEI 3005643748), from 2010 to 2015. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Mar 14, 2014. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2010–2015
- 1 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch Monitoring5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
No financial disclosure info in investigator agreement
A signed agreement was not obtained from each participating investigator that includes sufficient accurate financial disclosure information to allow the sponsor to submit a complete and accurate certification or disclosure statement.
Sponsor records re QSR/GMP & non-significant risk studies
For an investigation subject to the abbreviated requirements, records that include a statement of the extent to which the regulations covering current good manufacturing practice would be followed in manufacturing the device reports are not all accurate, complete, and current.
Abbreviated requirements / non-significant risk study
A clinical investigation that was determined to be a non-significant risk device study does not meet the abbreviated requirements for investigational device exemptions (IDEs).
Sponsors' general responsibilities
For an investigational study, reviewing IRBs were not promptly informed of significant new information about an investigation.
No IRB approval for treatment use
A treatment use study of an investigational device did not comply with the regulations governing informed consent.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Rationale documented for no investigation
Records for complaints where no investigation was made do not include required information.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training records
Personnel training is not documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
Show 1 more citation
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained, and implemented.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 812.2Applicability1 of 3
- 21 CFR 812.40General responsibilities of sponsors1 of 3
- 21 CFR 812.43Selecting investigators and monitors1 of 3
- 21 CFR 812.140Records1 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.22Quality audit1 of 3
Show all 12
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.198Complaint files1 of 3
- 21 CFR 812.36Treatment use of an investigational device1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005643748 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005643748. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.