FDA has 7 inspections on file for Dr. Reddy's Laboratories Limited in Nalgonda, India (FEI 3005447965), from 2009 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.

The most recent OAI inspection ended on Jan 31, 2015. FDA issued a warning letter on (CDER). No court action or import refusal is filed under this FEI.

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for API manufacturing, packing and analysis, through Dec 31, 2027. Its registrant, Dr. Reddy's Laboratories Limited (D-U-N-S 650562841), has registered 13 other sites.

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2009–2025
1 OAI · 6 VAI · 0 NAI
Inspections with 483 observations shown
2 of 7
Form 483s posted by FDA
↗
2 posted
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 3 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Reprocessing controls for APIs

      Production system 1. Your system for reprocessing of APIs and API intermediates a) For example, you do not adequately identify reprocess with a unique material number or provide an identifier manufactured using reprocessed material from (b)<) incident. b) You do not adequately place (b)<) API re reprocessed lots on appropriate stability monitoring to the life cycle of the validated procedure by ensuring (with stability testing) reprocessed APIs and API intermediates (post PV) in accordance with USP monograph criteria and throughout using reprocessed materials. c) Your procedure for skip testing allows skip testing of intermediate (b)(4) which have been previously rejected and reprocessed. Yet appropriate justification why reduced testing :should batches which have initially been rejected due to failure

    2. Observation 2

      Reference standard retest periods

      Laboratory System 2. Retest periods for references standards used in the notes established with supporting analytical data. You have not appropriately validated the effective shelf lives suppliers, which are used for QC testing of (b)<). The period of (b)<) for third-party supplied API and A standard op and no validation studies have been performed.

    Show observation 3
    1. Observation 3

      Qualification of HPLC columns

      3. Failure to maintain laboratory equipment used in the analytical that the equipment is suitable for use in the execution. You have not qualified your HPLC columns used for that the column performance is appropriate and suitable to perform the specified USP monograph method for chromatographic purity related substances. Per your firm you created a second analytical method to adequate (b)<) peak... due to "poor column performance/ deg[illegible]" determine the number of injections which the column suitability and must be discarded/replaced.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  4. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssurancePosted 483 · 3 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Test method specificity not established

      The specificity of test methods has not been established. Specifically, A. Your firm failed to perform the stress (forced degradation) studies for the ®® USP non-compendia optical purity test by HPLC analytical test method during the test method validation. The validation number 20 EAiVIVA 002, dated 06/26/2002 does not contain the specificity/selectivity studies necessary to demonstrate the suitability and adequacy of the analytical test method and specifications, S-08-®@-USP/13, used for the release and stability testing of this Active Pharmaceutical Ingredient (API). In addition, the current HPLC chromatographic purity test profile was found different than that obtained during the method validation. The retention time for [illegible] peak changed from about [illegible] minutes in the validation to [illegible] minutes during the routine testing. B. The data supporting validation activities performed by your firm to demonstrate the suitability and adequacy of the analytical method used for the release and stability testing of ®® USP Active Pharmaceutical Ingredient (APT) was found to be inadequate and incomplete. Data and documents presented by your firm supporting the validations number: 20-®® AVR-001-10 for the non-compendia Related Substances analytical [illegible] test method and specification, S-08-®@_USP/07, does not contain the specificity/selectivity studies necessary to determine if the method is a stability indicating method for the testing of the API. Validation documents showed that the validation of the methods does not contain complete forced degradation studies. Degradation of the main peak was not achieved. In addition, the routine analytical test method specification, S-08-®@_USP/07, failed to contain the precision standard system suitability (%RSD) that [illegible] the injection of the sample solution. Nevertheless, the validation of the test methods were approved by QCU.

    2. Observation 2

      Instrument calibration deficiencies

      Procedures describing the calibration of instruments are deficiently written or followed. Specifically, a) The calibration of the Gas Chromatographic (GC) instruments was incomplete. The review of the ®@ Operational Calibration for the Agilent GC/Headspace (GC/HS) QC-206, conducted by the outside contractor and as in-house calibration, did not include the HS oven temperature, noise and drift, signal to noise and detector accuracy tests as part of the GC/HS calibration. The GC instruments were used for the Residual Solvents test determination for release of the APIs and raw materials. b) The micro balance, QC-465, was used outside its calibration range (usage range: 10 mg to 2 g) for the testing of ®@ USP Active Pharmaceutical Ingredient (API) during release and stability studies. The firm performed and documented the weight of [illegible] mg, [illegible] mg, [illegible] mg, [illegible] mg and [illegible] mg for the standards of the related substance test for ®® USP, batch number [illegible] using the balance number QC-465 that was certified for weights in the range of 10 mg to 2 g. In addition, the linearity of this balance was performed with three points starting with 10 mg and the accuracy with 10 mg that do not support the use of the balance below 10 mg.

    Show observation 3
    1. Observation 3

      Unvalidated USP method modifications

      Records maintained of any modification of an established method employed in testing do not include the reason for the modification, the data to verify that the modification produced results that are at least as accurate and reliable for the material being tested as the established method. Specifically, Your firm failed to follow the USP monograph identification test method for the Ultraviolet Absorption <197U> that defined how to perform the test and calculate the results for ®® API's. For [illegible] it was noted that the absorbance readings were obtained at a maximum wavelength other than [illegible] nm as prescribed by the USP test monograph. During the review of ®® batches [illegible] it was observed that the absorbance was determined at approximately [illegible] nm instead of the exact wavelength of [illegible] nm as established by the USP test monograph. Also, the final calculations were performed with the absorbance readings instead of absorptivity as prescribed by the USP monograph. These modifications to the USP test method and calculation formulas were not validated or the equivalencies demonstrated.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssuranceCitations not published
  6. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  7. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDER · DrugsFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

FDA drug establishment registration

FDA registration search
Registered operations
API manufacturing, packing, analysis
Registration runs through
Dec 31, 2027
Registrant
Dr. Reddy's Laboratories LimitedD-U-N-S 650562841

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Other sites with the same registrant

FDA’s registration file names Dr. Reddy's Laboratories Limited (D-U-N-S 650562841) as the registrant for this site and 13 others. Listed: the 12 with the most warning letters, then the most recent inspections. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. This site’s records are filed under FEI 3005447965 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every inspection and enforcement record on this page is filed under FEI 3005447965. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.