IDEV Technologies, Inc.
253 W Medical Center Blvd, Webster, Texas, United States
FEI 3005325609
FDA has 5 inspections on file for IDEV Technologies, Inc. in Webster, Texas (FEI 3005325609), from 2009 to 2013. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.
The most recent OAI inspection ended on Jan 21, 2009. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2009–2013
- 2 OAI · 1 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complete files maintained
Complete complaint files are not maintained.
Nonconforming product, Lack of or inadequate procedures
Procedures were not complete and followed for the control of products that do not conform to specifications.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been fully validated and approved according to established procedures.
Documentation - review in response to changes or deviations
There is no documentation of the review and evaluation of a process and revalidation of a process conducted in response to changes or process deviations.
Documentation
Process validation activities and results have not been fully documented.
Show 3 more citations
Results of investigation
Records of complaint investigations do not include the dates and results of the investigation.
Documentation
Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities and the actions needed to correct or prevent recurrence of nonconforming product and other quality problems.
Change records, content
Records of changes did not include a description of the change and identification of the affected documents.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Incomplete, ambiguous, conflicting design requirements
Design input requirements that are incomplete and ambiguous were not addressed.
Design validation - user needs and intended uses
Design validation did not ensure that devices conform to defined user/patient needs and intended uses.
Design verification - unresolved discrepancies
Unresolved discrepancies were noted at the completion of the design verification.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.198Complaint files1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005325609 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005325609. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.