Nano-Ditech Corporation
259 Prospect Plains Rd Bldg K Ste 100, Cranbury, New Jersey, United States
FEI 3005174594
FDA has 6 inspections on file for Nano-Ditech Corporation in Cranbury, New Jersey (FEI 3005174594), from 2009 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.40.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2009–2026
- 0 OAI · 1 VAI · 5 NAI
- Inspections with 483 observations shown
- 2 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of procedures, or not maintained
Document control procedures were not implemented.
Equipment control activity documentation
Equipment inspections and maintenance activities have not been documented.
Verification vs validation
There is no corrective and preventive action procedure to determine when verification can be conducted in lieu of validation.
Records, DHR {see also 820.184(e)}
Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented and kept in the DHR.
DMR production process specifications
The device master record does not include or refer to the location of all production and process specifications.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.40[Reserved]2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.181Device master record2 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.120Device labeling1 of 2
- 21 CFR 820.184Device history record1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005174594 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005174594. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.