G3 Medical, Inc. dba G3 Packaging
81 Thompson St, Asheville, North Carolina, United States
FEI 3004726911
FDA has 2 inspections on file for G3 Medical, Inc. dba G3 Packaging in Asheville, North Carolina (FEI 3004726911), from 2009 to 2014. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 15 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.25(b), 21 CFR 820.75(a) and 21 CFR 820.75(c).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2009–2014
- 2 OAI · 0 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation activities and results have not been adequately documented.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Show 10 more citations
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Equipment Installation, Placement, Specified Requirements
The equipment used in the manufacturing process does not meet specified requirements.
Equipment control activity documentation
Equipment maintenance activities have not been documented.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Documentation of validated process performance
There is inadequate documentation of monitoring and control methods and data for a validated process.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Documentation - review in response to changes or deviations
There is no documentation of the revalidation of a process performed in response to changes or process deviations.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control instructions
Process controls are not conducted in accordance with documented instructions and standard operating procedures.
Process changes - review, evaluation and revalidation
A validated process was not reviewed and evaluated when changes or process deviations occurred.
Documentation
Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities, the verification or validation of corrective actions, and dissemination of information about quality problems or nonconforming product to responsible parties.
Acceptance records
The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately and fully validated and approved according to established procedures.
Show 2 more citations
Training
Employees have not been adequately trained.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures were not complete and implemented for monitoring and control of process parameters for validated processes.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.25Personnel2 of 2
- 21 CFR 820.70Production and process controls2 of 2
- 21 CFR 820.75Process validation2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.184Device history record2 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004726911 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004726911. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.