Innovasis, Inc
614 E 3900 S, Salt Lake City, Utah, United States
FEI 3004719693
FDA has 5 inspections on file for Innovasis, Inc in Salt Lake City, Utah (FEI 3004719693), from 2009 to 2019. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.30(g).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2009–2019
- 0 OAI · 2 VAI · 3 NAI
- Inspections with 483 observations shown
- 2 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures to meet core CTGP
Procedures appropriate to meet core CGTP requirements for all steps that you perform in the manufacture of HCT/Ps were not established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Timely informing of consignees
Consignees were not informed in writing at or before the time of distribution of an HCT/P, of the requirements related to HCT/P tracking.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 1 more citation
Storage temperatures recorded, maintained
Storage temperatures of HCT/Ps were not recorded.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design control - no procedures
Procedures to control the design process of the device were not implemented.
Design changes - Lack of or Inadequate Procedures
Procedures were not established for the identification, documentation, validation or verification, review, and approval of design changes before their implementation.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Design validation- Lack of or inadequate procedures
Procedures for validating the device design were not complete.
Processing MDRs {see also 803, 804}
Complaint handling procedures have not been implemented to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.
Show 7 more citations
Acceptance criteria and essential outputs
Design output procedures do not ensure that essential design outputs are identified.
Design input procedures - addressing conflicting requirement
Procedures for addressing incomplete, ambiguous, or conflicting design input requirements were not established.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Analysis of data/reports from data sources
Not all data from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated and approved according to established procedures.
Design verification - Lack of or inadequate procedures
Design verification did not confirm that the design output meets the design input requirements.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.198Complaint files1 of 2
- 21 CFR 1271.180Procedures1 of 2
- 21 CFR 1271.260Storage1 of 2
- 21 CFR 1271.290Tracking1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004719693 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004719693. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.