Biometerplususa Inc.
48531 Warm Springs Blvd Ste 416, Fremont, California, United States
FEI 3004525092
FDA has 5 inspections on file for Biometerplususa Inc. in Fremont, California (FEI 3004525092), from 2009 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 5 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2009–2026
- 0 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Jun 16, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 4.1.5
Inadequate monitoring and control over outsourced processes
Outsourced processes that affect product conformity to requirements were not monitored and controlled.
- ISO 13485:2016 Clause 8.2.2
Complaint handling procedures lack required elements
Procedures for timely complaint handling do not include requirements and responsibilities for determining the need to report the information to FDA.
- ISO 13485:2016 Clause 7.5.8
No documented system to assign unique device identification (UDI)
A system to assign unique device identification (UDI) to the medical device was not adequately documented.
- ISO 13485:2016 Clause 4.1.2
Risk-based approach not applied to QMS processes
A risk-based approach to the control of the appropriate processes needed for the quality management system has not been applied.
- ISO 13485:2016 Clause 7.6(a)
Measuring equipment not adequately calibrated or verified
Measuring equipment was not calibrated at specified intervals or prior to use.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training records
Personnel training is not documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design control - no procedures
Procedures to control the design process of the device were not complete.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not complete.
Management review - Lack of or inadequate procedures
Procedures for management review are not complete.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure that equipment is routinely maintained were not complete.
Show 3 more citations
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been completed.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not complete.
Sampling plans
Sampling plans are not based on valid statistical rationale.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Jun 16, 2026. Each one asked for a Form 483.
Inspection ending Mar 25, 2026VAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on May 4, 2026. FDAgent holds no copy that FDA posted.
- Jun 16, 2026FDA-FOIA-2026-5845Form 483· 1 of 4 sitesRecords released
- May 6, 2026FDA-FOIA-2026-4586Form 483· 1 of 5 sitesRecords released
- Apr 7, 2026FDA-FOIA-2026-3441Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
- 21 CFR 820.198Complaint files1 of 3
Show all 9
- 21 CFR 820.250Statistical techniques1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004525092 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004525092. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.