Cardiovascular Imaging Technologies, LLC.
4320 Wornall Rd Ste 114, Kansas City, Missouri, United States
FEI 3004504920
FDA has 4 inspections on file for Cardiovascular Imaging Technologies, LLC. in Kansas City, Missouri (FEI 3004504920), from 2009 to 2024. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2009–2024
- 0 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - software validation documentation
Results of the validation of the device software were not adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy satisfies the requirements of the quality system regulations.
Management review - defined interval, sufficient frequency
Management reviews were not conducted at defined intervals and with sufficient frequency.
Lack of or inadequate instructions
Installation, inspection, and test procedures have not been documented such that the device will perform as intended after installation.
Corrective action taken
Records of complaint investigations do not include any corrective action taken.
Rationale documented for no investigation
Records for complaints where no investigation was made do not include the reason why and the name of the individual responsible for the decision not to investigate.
Show 4 more citations
Documentation
Corrective and preventive action activities have not been documented, including the verification or validation of corrective actions.
Identification procedures, Lack of or inadequate procedures
Procedures for identifying product throughout all stages of receipt, production, packaging, distribution, and installation were not documented.
Quality Audit/Reaudit - conducted
Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.
Training records
Employee training is not fully documented.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.25Personnel1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.60Identification1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.170Installation1 of 2
- 21 CFR 820.198Complaint files1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004504920 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004504920. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.