Nuga Medical Co., Ltd.
185 Jiraeul-Ro, Jijeong-Myeon, Wonju, Korea (the Republic of)
FEI 3004006265
FDA has 3 inspections on file for Nuga Medical Co., Ltd. in Wonju, Korea (the Republic of) (FEI 3004006265), from 2014 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Aug 28, 2014. 9 import refusals are filed under this FEI, the latest on . FDA’s Import Alert 89-08 Red List names this firm at this address. FDA printed no FEI on the entry. No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2014–2022
- 1 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 9 refusals
- Import alerts
- 89-08Red List
- Matched on FDA’s firm name and street address
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Records of complaint investigation
Records of complaint investigations do not include required information.
Identification procedures, Lack of or inadequate procedures
Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Equipment control activity documentation
Equipment calibrations and maintenance activities have not been documented.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a specification and procedure have not been established.
Documentation
Corrective and preventive action activities and/or results have not been documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.60Identification1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.198Complaint files1 of 2
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3004006265, one row per refused product line.
9 refusals from 2018-04-02 to 2021-04-29 · charged DEVGMPS on 7, REGISTERED on 6 · no FDA sample analysis
FDA firm profile- 2021-04-29
MASSAGER, THERAPEUTIC, ELECTRIC89ISA
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-4617905-9/11/1
- 2019-09-10
MASSAGER, THERAPEUTIC, ELECTRIC89ISA
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-2964640-2/11/1
- 2018-04-16
PAD, HEATING, POWERED89IRT
DEVGMPS 520(f); 801(a)(1)
Division of West Coast ImportsBFH-0801079-2/11/1
- 2018-04-02
PILLOW, CERVICAL (FOR MILD SLEEP APNEA)77MYB
REGISTERED 502(o), 801(a)(3); MISBRANDING
DEVGMPS 520(f); 801(a)(1)
Division of West Coast ImportsBFH-0800811-9/11/1F
- 2018-04-02
PAD, HEATING, POWERED89IRT
REGISTERED 502(o), 801(a)(3); MISBRANDING
DEVGMPS 520(f); 801(a)(1)
Division of West Coast ImportsBFH-0800811-9/11/1E
Show 4 more
- 2018-04-02
PAD, HEATING, POWERED89IRT
REGISTERED 502(o), 801(a)(3); MISBRANDING
DEVGMPS 520(f); 801(a)(1)
Division of West Coast ImportsBFH-0800811-9/11/1D
- 2018-04-02
PAD, HEATING, POWERED89IRT
REGISTERED 502(o), 801(a)(3); MISBRANDING
DEVGMPS 520(f); 801(a)(1)
Division of West Coast ImportsBFH-0800811-9/11/1C
- 2018-04-02
PAD, HEATING, POWERED89IRT
REGISTERED 502(o), 801(a)(3); MISBRANDING
DEVGMPS 520(f); 801(a)(1)
Division of West Coast ImportsBFH-0800811-9/11/1B
- 2018-04-02
MASSAGER, THERAPEUTIC, ELECTRIC89ISA
REGISTERED 502(o), 801(a)(3); MISBRANDING
DEVGMPS 520(f); 801(a)(1)
Division of West Coast ImportsBFH-0800811-9/11/1A
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov
Date published 2009-09-16
- FDA’s entry
- Nuga Medical Co., Ltd. · 185 Jiraeul-Ro, Jijeong-Myeon, Wonju, Gangwon KOREA (THE REPUBLIC OF)
- FDA’s inspection record for FEI 3004006265
- Nuga Medical Co., Ltd. · 185 Jiraeul-Ro, Jijeong-Myeon, Wonju, Korea (the Republic of)
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
5 products
- 84 G - - ZJStimulator, Nerve, Transcutaneous, For Pain Relief
- 89 I - - PFStimulator, Muscle, Powered
- 89 I - - SAMassager, Therapeutic, Electric
Show 2 more products
- 89 M - - PHStimulator, Electrical, Muscle
- 89 N - - GXStimulator, Muscle, Powered, for Muscle Conditioning
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004006265 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004006265. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.