Skip to content

Event Medical Ltd

Glenrock Business Park, Unit 11; 29 Glenrock Business Park, Galway, Ireland

FEI 3003638180

FDA has one inspection on file for Event Medical Ltd in Galway, Ireland (FEI 3003638180). It ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 10 citations from this inspection: 8 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2010
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality plan

      A quality plan was not implemented.

    2. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    3. Design validation unresolved discrepancies

      Unresolved discrepancies were noted at the completion of the design validation.

    4. Quality policy and objectives

      The quality policy and objectives were not complete.

    5. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    Show 5 more citations
    1. Procedures

      The procedures addressing documentation of corrective and preventive action activities were not defined.

    2. Documentation

      Corrective and preventive action activities have not been documented, including analysis of sources of quality data, investigations of causes of nonconformities, the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, and the verification or validation of corrective actions.

    3. Identification of actions needed

      Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.

    4. Quality Audit/Reaudit - conducted

      Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.

    5. Document changes, review and approval, communication

      Changes to documents were not reviewed and approved by an individual in the same function or organization that performed the original review.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003638180 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003638180. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.