Bieffe Medital SpA
Via Stelvio 94, Sondalo, Italy
FEI 3003591774DUNS 437668413
Company: Vantive US Healthcare LLC · Linked by exact D-U-N-S match (119533513) in FDA registration data
FDA has 6 inspections on file for Bieffe Medital SpA in Sondalo, Italy (FEI 3003591774), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.30(g).
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing, labeling, analysis and sterilization, through Dec 31, 2026. Its registrant, Vantive US Healthcare LLC (D-U-N-S 119533513), has registered 3 other sites.
Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2009–2023
- 0 OAI · 4 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Aug 12, 2021
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a process and procedure have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design history file
The design history file was not established or maintained.
Changes to correct/prevent quality problems
Changes in methods and procedures needed to correct and prevent identified quality problems are not implemented and recorded.
CAPA verification/validation of corrective/preventive action
Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.
Investigation
Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Show 2 more citations
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated where necessary.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Aug 12, 2021. It asked for a Form 483.
Inspection ending Jun 28, 2019VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Aug 12, 2021FDA-FOIA-2021-5280Form 483Withdrawn
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.198Complaint files1 of 2
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, packing, labeling, analysis, sterilization
- Registration runs through
- Dec 31, 2026
- Registrant
- Vantive US Healthcare LLCD-U-N-S 119533513
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Other sites with the same registrant
FDA’s registration file names Vantive US Healthcare LLC (D-U-N-S 119533513) as the registrant for this site and the 3 others below. That exact D-U-N-S number is the only link. A registrant is the firm that filed the registration, which is not always the owner.
- Vantive LimitedLiverpool, United Kingdom · FEI 10003586971 inspection1latestJun 17, 2016VAINone on file
- VANTIVE MANUFACTURING LIMITEDCounty Mayo F23 XR, Ireland · FEI 0000000405Registered; no FDA inspection on file
- Baxter, S.A. de C.V.Morelos, Mexico · FEI 0000006898Registered; no FDA inspection on file
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. This site’s records are filed under FEI 3003591774 in FDA’s data; other sites under its registrant are listed in summary only. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every inspection and enforcement record on this page is filed under FEI 3003591774. Other sites filed under the same registrant D-U-N-S are listed in summary only. Records filed under another FEI or name are otherwise not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.