Skip to content

Westridge Laboratories Inc

1671 E Saint Andrew Pl, Santa Ana, California, United States

FEI 3003169402

FDA has 4 inspections on file for Westridge Laboratories Inc in Santa Ana, California (FEI 3003169402), from 2013 to 2023. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below. Two FOIA requests asked FDA for the 483 from a later inspection, ending Dec 29, 2025, that FDA's Data Dashboard doesn't list yet.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2013–2023
0 OAI · 1 VAI · 3 NAI
Inspections with 483 observations shown
2 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Feb 23, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Rationale documented for no investigation

      Records for complaints where no investigation was made do not include required information.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been established.

    2. Supplier oversight

      The type and extent of control to be exercised over the product, services, suppliers, and contractors was not clearly defined.

    3. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    4. Design validation - documentation

      The results of design validation, including identification of the design, method(s), and the individual(s) performing validation, were not adequately documented in the design history file.

    5. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a specification have not been established.

    Show 3 more citations
    1. Investigation of device failures

      Complaints involving the possible failure of a device and packaging to meet any of its specifications were not evaluated and investigated where necessary.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Feb 23, 2026. Each one asked for a Form 483.

  • Inspection ending Dec 29, 2025

    2 requests. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Mar 6, 2026. FDAgent holds no copy that FDA posted.

    1. Feb 23, 2026FDA-FOIA-2026-1859Form 483Records released
    2. Feb 12, 2026FDA-FOIA-2026-1494Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 5 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003169402 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003169402. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.