IND Diagnostic Inc.
1629 Foster's Way, Delta, Canada
FEI 3003148151
FDA has one inspection on file for IND Diagnostic Inc. in Delta, Canada (FEI 3003148151). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 9 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
FDA’s Import Alert 89-04 Red List names this firm at this address. FDA printed no FEI on the entry. No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2010
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- 89-04Red List
- Matched on FDA’s firm name and street address
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quantity manufactured
The device history record does not include the quantity of devices manufactured.
Lack of or inadequate final acceptance procedures
Procedures for acceptance or rejection of finished device production runs, lots, or batches were not defined.
Design transfer - Lack of or inadequate procedures
Procedures were not established to ensure that the device design was correctly transferred into production.
Dissemination of problem information - procedure
The corrective and preventive action procedures addressing dissemination of information on quality problems and nonconforming product to responsible individuals were not established.
Documentation
Corrective and preventive action activities have not been documented, including analysis of sources of quality data, the verification or validation of corrective actions, and dissemination of information about quality problems or nonconforming product to responsible parties.
Show 4 more citations
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been fully validated and approved according to established procedures.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product were not documented.
Review of data and documentation
Finished devices were released for distribution before the associated data and documentation was reviewed.
ID label, labeling
The device history record does not include the primary identification label and labeling for each device.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.75Process validation1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.90Nonconforming product1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.184Device history record1 of 1
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov
Date published 2013-02-25
- FDA’s entry
- IND Diagnostic Inc. · 1629 Foster's Way, Delta, British Columbia CANADA
- FDA’s inspection record for FEI 3003148151
- IND Diagnostic Inc. · 1629 Foster's Way, Delta, Canada
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
17 products
- 73 - - - --Anesthesiology
- 74 - - - --Cardiovascular
- 76 - - - --Dental
Show 14 more products
- 77 - - - --Ear,Nose And Throat
- 78 - - - --Gastroenterological & Urological
- 79 - - - --General & Plastic Surgery
- 80 - - - --General Hospital/Personal Use
- 84 - - - --Neurological
- 85 - - - --Obstetrical & Gynecological
- 86 - - - --Ophthalmic
- 87 - - - --Orthopedic
- 89 - - - --Physical Medicine
- 90 - - - --Radiology
- 94 - - - --Ionizing Non-Medical Devices and Components
- 95 - - - --Light Emitting Non-Device Products
- 96 - - - --Radio Frequency Emitting Products
- 97 - - - --Sound Emitting Products
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003148151 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003148151. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.