Innolatex Sdn. Bhd.
Lot 591 & 594, Pandamaran Industrial Estate, Persiaran Raja Lumu, Port Klang, Malaysia
FEI 3003117682
FDA has 5 inspections on file for Innolatex Sdn. Bhd. in Port Klang, Malaysia (FEI 3003117682), from 2011 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Jan 20, 2011. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.75(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2011–2026
- 1 OAI · 1 VAI · 3 NAI
- Inspections with 483 observations shown
- 2 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process changes - review, evaluation and revalidation
A validated process was not revalidated when changes or process deviations occurred.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
Examination for accuracy
Labeling was not sufficiently examined by a designated individual for accuracy including control numbers before release.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Production processes
Production processes were not controlled to ensure that a device conforms to its specifications.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design history file
The design history file was not established and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Design control - no procedures
Procedures for design control have not been established.
Show 1 more citation
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.75Process validation2 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.120Device labeling1 of 2
- 21 CFR 820.184Device history record1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003117682 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003117682. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.