GS Medical Packaging Inc
82 Akron Rd Unit 2, Etobicoke, Canada
FEI 3003067378
FDA has 3 inspections on file for GS Medical Packaging Inc in Etobicoke, Canada (FEI 3003067378), from 2013 to 2022. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Jan 31, 2013. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2013–2022
- 1 OAI · 0 VAI · 2 NAI
- Inspections with 483 observations shown
- 1 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed and implemented.
Show 2 more citations
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 3 inspections.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.184Device history record1 of 1
- 21 CFR 820.198Complaint files1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003067378 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003067378. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.