Skip to content

Medical & Clinical Consortium Inc

6880 Oran Cir Ste G, Buena Park, California, United States

FEI 3003043277

FDA has 2 inspections on file for Medical & Clinical Consortium Inc in Buena Park, California (FEI 3003043277), from 2009 to 2010. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
2, 2009–2010
0 OAI · 0 VAI · 2 NAI
Inspections with 483 observations shown
1 of 2
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices13 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by consultants were not established.

    2. Documentation

      Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities, the verification or validation of corrective actions, implementation of corrective and preventive actions, and dissemination of information about quality problems or nonconforming product to responsible parties.

    3. Management representative

      No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.

    4. Quality Audit/Reaudit - documentation

      The dates of quality audits have not been documented.

    5. Design history file

      The design history file was not established or maintained.

    Show 8 more citations
    1. Training

      Employees have not been adequately trained.

    2. Design transfer - Lack of or inadequate procedures

      Procedures were not implemented to ensure that the device design was correctly transferred into production.

    3. Design verification - Lack of or inadequate procedures

      Design verification did not confirm that the design output meets the design input requirements.

    4. Design control - no procedures

      Procedures to control the design process of the device were not implemented.

    5. Management review - Lack of or inadequate procedures

      Procedures for management review are not defined.

    6. Lack of procedures, or not maintained

      Document control procedures were not established.

    7. Equipment suitability & capability

      Certain measuring equipment is not suitable for its intended purposes or capable of producing valid results.

    8. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure that equipment is routinely calibrated, inspected, checked, and maintained were not established.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003043277 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003043277. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.