LaserSight Technologies, Inc.
931 S. Semoran Boulevard, Winter Park, Florida, United States
FEI 3002741596
FDA has 3 inspections on file for LaserSight Technologies, Inc. in Winter Park, Florida (FEI 3002741596), from 2009 to 2012. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.198(a) and 21 CFR 820.22.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2009–2012
- 0 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
No multiple safety interlock
A laser product was not provided with either multiple safety interlocks or a means to preclude removal or displacement of the interlocked portion of the protective housing, when failure of a single interlock allowed human access to a level of laser radiation in excess of the accessible emission limits of Class IIIa.
Remote interlock connector
Each laser system classified as a Class IIIb or Class IV laser product had no means to prevent, when the terminals of the connector are not electrically joined, human access to all laser and collateral radiation from the laser product in excess of the accessible emission limits of Class I and table VI.
User information, adequate instructions
Adequate instructions for each laser product were not provided or caused to be provided for assembly, operation, maintenance, and clear warnings concerning precautions to avoid possible exposure to laser and collateral radiation in excess of the accessible emission limits specified in the regulations.
Reproduction of affixed information
The manufacturer did not provide or cause to be provided in all catalogs, specification sheets, and descriptive brochures pertaining to each laser product, a legible reproduction of the class designation, the warning, and the information required for positions 1, 2, and 3 of the applicable logotype required b to be affixed to the product.
Surveying, leveling, alignment laser products
Each surveying, leveling, and alignment laser product does not comply with all of the requirements for a Class I, IIa, II, and IIIa laser product and does not prevent human access to laser radiation in excess of the accessible emission limits of Class IIIa.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedure for analysis of data sources
The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not defined.
Production processes
Production processes were not complete and controlled to ensure that a device conforms to its specifications.
Sufficient personnel
Insufficient personnel to assure that all procedures are appropriately carried out as required by the quality system.
Quality Audits - defined intervals
Quality audits did not verify that the quality system is effective in fulfilling your quality system objectives.
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been defined.
Show 2 more citations
Foreign complaint files
Complaint files are not reasonably accessible in the United States at a location where the manufacturer's records are regularly kept.
Analysis of data/reports from data sources
Not all data and reports from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 1040.10Laser products1 of 3
- 21 CFR 1040.11Specific purpose laser products1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002741596 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002741596. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.