Wassenburg Medical, Inc.

144 Railroad Dr, Ivyland, Pennsylvania, United States

FEI 2523209

FDA has 7 inspections on file for Wassenburg Medical, Inc. in Ivyland, Pennsylvania (FEI 2523209), from 2010 to 2018. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Apr 24, 2015. No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.75(a) and 21 CFR 820.30(f).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
7, 2010–2018
3 OAI · 1 VAI · 3 NAI
Inspections with 483 observations shown
3 of 7
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Acceptance records

      The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.

    2. Environmental control Lack of or inadequate procedures

      Appropriate procedures have not been defined and documented for controlling environmental conditions.

    3. Not approved or obsolete document retrieval

      Documents that were not approved were observed at a location where they are being used.

    4. Design verification acceptance criteria

      Acceptance criteria were not documented prior to the performance of verification activities.

    5. Design transfer - Lack of or inadequate procedures

      Procedures were not completed to ensure that the device design was correctly transferred into production.

    Show 3 more citations
    1. Lack of or inadequate In-process acceptance procedures

      Acceptance procedures to ensure that specified requirements for in-process product are met were not defined.

    2. Management responsibility

      Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated and approved according to established procedures.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Management ensuring quality policy is understood

      Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.

    2. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    3. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    4. Equipment Installation, Placement, Specified Requirements

      The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured.

    5. Design validation - documentation

      The results of design validation, including identification of the design, were not adequately documented in the design history file.

    Show 4 more citations
    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    2. Records of MDR Investigation

      Investigation records of MDR reportable complaints do not include required information.

    3. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    4. Complaints

      Complaint files are not adequately maintained.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 7 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2523209 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2523209. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.