DeVilbiss Healthcare LLC
100 Devilbiss Dr, Somerset, Pennsylvania, United States
FEI 2515872
FDA has 7 inspections on file for DeVilbiss Healthcare LLC in Somerset, Pennsylvania (FEI 2515872), from 2009 to 2024. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Aug 17, 2022. FDA issued a warning letter on (CDRH · Office of Medical Device and Radiological Health Operations (Division 1)). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.198(a) and 21 CFR 820.30(f).
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 7, 2009–2024
- 2 OAI · 0 VAI · 5 NAI
- Inspections with 483 observations shown
- 2 of 7
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest FY2023
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Records of complaint investigation
Records of complaint investigations do not include required information.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 4 more citations
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been established.
Rationale documented for no investigation
Records for complaints where no investigation was made do not include required information.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Aug 17, 2023
1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Mar 9, 2026. FDAgent holds no copy that FDA posted.
- Feb 27, 2026FDA-FOIA-2026-2029Form 483Records released
Inspection ending Aug 25, 2021OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-6443Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · Office of Medical Device and Radiological Health Operations (Division 1) · DevicesLetter on fda.govFDA record
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2515872, one row per refused product line.
1 refusal on 2023-02-07 · charged NOT LISTED on 1, RXPERSONAL on 1 · no FDA sample analysis
FDA firm profile- 2023-02-07
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)73BZD
NOT LISTED 502(o), 801(a)(3); MISBRANDING
RXPERSONAL 502(a), 502(f)(1), 801(a)(3), MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-3061528-1/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2515872 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2515872. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.