Ambulatory Monitoring Inc.
731 Saw Mill River Rd, Ardsley, New York, United States
FEI 2430863
FDA has 5 inspections on file for Ambulatory Monitoring Inc. in Ardsley, New York (FEI 2430863), from 2010 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection: 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2010–2024
- 0 OAI · 4 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Maintained
Records of complaint investigations are not maintained by the formally designated unit.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design output - documentation
Design output was not adequately documented before release.
Acceptable supplier records, inadequate records
Records of acceptable suppliers have not been adequately established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Uniform and timely processing
Complaint handling procedures have not been implemented to ensure that all complaints are processed in a uniform and timely manner.
Documentation
Acceptance activities were not fully documented.
Quality Audit/Reaudit - conducted
Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2430863 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2430863. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.