Skip to content

Medicia Holdings LLC

2351 US Highway 130, Dayton, New Jersey, United States

FEI 2243280

FDA has 2 inspections on file for Medicia Holdings LLC in Dayton, New Jersey (FEI 2243280), from 2010 to 2011. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 6 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2010–2011
0 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Calibration/Inspection/Checking not done

      Routine calibration, inspection, and checking of automatic equipment is not performed according to a written program designed to assure proper performance.

    2. Written records kept in individual logs

      Written records of major equipment use are not included in individual equipment logs.

    3. Backup data not assured as exact and complete

      Backup data is not assured as secure from alteration, erasure or loss through keeping hard copy or alternate systems.

    4. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not fully followed.

    5. Written procedures fail to include

      Written procedures for cleaning and maintenance fail to include maintenance and cleaning schedules, description in sufficient detail of methods, equipment and materials used, and instructions for protection of clean equipment from contamination prior to use.

    Show 1 more citation
    1. Complaints reviewed by Quality Control Unit

      Written procedures describing the handling of complaints do not include provisions for a determination as to the need for an investigation of any unexplained discrepancy.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design control - no procedures

      Procedures to control the design process of the device were not established.

    2. Investigation

      Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.

    3. Procedures

      The procedures addressing documentation of corrective and preventive action activities were not complete.

    4. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not complete.

    5. Analysis of data/reports from data sources

      Not all data from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2243280 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2243280. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.