Cetylite Industries, Inc.
9051 River Rd, Pennsauken, New Jersey, United States
FEI 2242327DUNS 001283704
FDA has 9 inspections on file for Cetylite Industries, Inc. in Pennsauken, New Jersey (FEI 2242327), from 2008 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the five FOIA requests for this site have asked FDA for this inspection's 483.
The most recent OAI inspection ended on Mar 2, 2023. FDA issued a warning letter on (CDER · Division of Pharmaceutical Quality Operations I). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing, labeling and analysis, through Dec 31, 2027. Its registrant is Cetylite Industries, Inc. (D-U-N-S 001283704).
Sources:FDA firm profile ↗Warning letter ↗FDA drug establishment registrations ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 9, 2008–2025
- 2 OAI · 4 VAI · 3 NAI
- Inspections with 483 observations shown
- 7 of 9
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 5 requests, latest Oct 15, 2025
- Warning letters
- · CDER · Division of Pharmaceutical Quality Operations I ↗
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
- Drug establishment registration
- Through
Inspection history
9 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Absence of Written Procedures
Your firm failed to establish written procedures for production and process controls designed to assure that the drug products have the identity, strength, purity, and quality that they are purported or represented to possess.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Written procedures not established/followed
Written procedures are not established for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product.
Procedures for non-sterile drug products
Procedures designed to prevent objectionable microorganisms in drug products not required to be sterile are not established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
Microbiological testing
Each batch of drug product required to be free of objectionable microorganisms is not tested through appropriate laboratory testing.
Cleaning / Sanitizing / Maintenance
Equipment and utensils are not cleaned, maintained and sanitized at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards, sampling plans and test procedures designed to assure that components, in-process materials and drug products conform to appropriate standards of identity, strength, quality and purity.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications and standards designed to assure that components, labeling and drug products conform to appropriate standards of identity, strength, quality and purity.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy and the failure of a batch or any of its components to meet any of its specifications whether or not the batch has been already distributed.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Bioresearch Monitoring8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.
Inadequate number of personnel
The number of qualified personnel is inadequate to perform the manufacture, processing, packing and holding of each drug product.
Expiration date lacking
Drug products do not bear an expiration date determined by appropriate stability data to assure they meet applicable standards of identity, strength, quality and purity at the time of use.
Failure to develop written procedures
Written procedures have not been developed for the surveillance, receipt, evaluation and reporting to FDA of postmarketing adverse drug experiences.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, sampling plans and test procedures designed to assure that components and drug products conform to appropriate standards of identity, strength, quality and purity.
Show 3 more citations
Status of Each Lot Identified
Each lot of components was not appropriately identified as to its status in terms of being quarantined, approved or rejected.
Failing drug products not rejected
Drug products failing to meet established specifications are not rejected.
Late submission
Each post marketing 15-day Alert report was not submitted to FDA within 15 calendar days of initial receipt of the information.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
No written record of investigation
Written records are not always made of investigations into the failure of a batch or any of its components to meet specifications.
Reprocessing procedures not written or followed
Procedures prescribing a system for reprocessing batches to insure that the reprocessed batches will conform with all established standards, specifications, and characteristics are not written.
Failing drug products not rejected
Drug products failing to meet established specifications are not rejected.
Control procedures to monitor and validate performance
Control procedures are not established which validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
Retest of approved components/containers/closures
Approved components are not retested or reexamined as appropriate for identity, strength, quality and purity after storage for long periods with subsequent approval or rejection by the quality control unit.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigations of discrepancies, failures
There is a failure to thoroughly review any unexplained discrepancy whether or not the batch has been already distributed.
Deviations from laboratory control requirements
Deviations from written specifications are not justified.
Laboratory equipment calibration records
Laboratory records do not include complete records of the periodic calibration of laboratory instruments.
Valid stability test methods
The written stability program for drug products does not include reliable and meaningful test methods.
Extent of discrepancy, failure investigations
Investigations of a failure of a batch or any of its components to meet any of its specifications did not extend to other batches of the same drug product and other drug products that may have been associated with the specific failure or discrepancy.
- NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Five requests to FDA under the Freedom of Information Act named this site, most recently on Oct 15, 2025. Each one asked for a Form 483.
Inspection ending Apr 7, 2025NAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 29, 2025. FDAgent holds no copy that FDA posted.
- Oct 15, 2025FDA-FOIA-2025-12016Form 483· 1 of 5 sitesRecords released
- May 12, 2025FDA-FOIA-2025-4197Form 483· 1 of 5 sitesRecords released
Inspection ending Dec 9, 2024VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Apr 30, 2025FDA-FOIA-2025-3874Form 483· 1 of 4 sitesClosed
- Mar 13, 2025FDA-FOIA-2025-2341Form 483· 1 of 2 sitesClosed
Inspection ending Sep 24, 2021OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Mar 1, 2022FDA-FOIA-2022-1648Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 18 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 9 inspections, with how many of them cited each.
- 21 CFR 211.192Production record review5 of 7
- 21 CFR 211.160General requirements4 of 7
- 21 CFR 211.165Testing and release for distribution3 of 7
- 21 CFR 211.22Responsibilities of quality control unit2 of 7
- 21 CFR 211.67Equipment cleaning and maintenance2 of 7
- 21 CFR 211.25Personnel qualifications1 of 7
- 21 CFR 211.80General requirements1 of 7
- 21 CFR 211.87Retesting of approved components, drug product containers, and closures1 of 7
Show all 18
- 21 CFR 211.100Written procedures; deviations1 of 7
- 21 CFR 211.110Sampling and testing of in-process materials and drug products1 of 7
- 21 CFR 211.113Control of microbiological contamination1 of 7
- 21 CFR 211.115Reprocessing1 of 7
- 21 CFR 211.137Expiration dating1 of 7
- 21 CFR 211.166Stability testing1 of 7
- 21 CFR 211.194Laboratory records1 of 7
- 21 CFR 310.305Records and reports concerning adverse drug experiences on marketed prescription drugs for human use without approved new drug applications1 of 7
- 21 CFR 820.22Quality audit1 of 7
- 21 CFR 820.50Purchasing controls1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDER · Division of Pharmaceutical Quality Operations I · DrugsLetter on fda.govFDA record
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2242327, one row per refused product line.
1 refusal on 2023-03-22 · charged AGR RX on 1 · no FDA sample analysis
FDA firm profile- 2023-03-22
BENZOCAINE (ANESTHETIC)60QDQ56
AGR RX 801(d)(1),(2); IMPORTATION RESTRICTED
Division of West Coast Imports8JB-0093850-4/11/17
FDA drug establishment registration
FDA registration search- Registered operations
- Manufacturing, packing, labeling, analysis
- Registration runs through
- Dec 31, 2027
- Registrant
- Cetylite Industries, Inc.D-U-N-S 001283704
From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2242327 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2242327. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.