Medtronic, Inc.
8200 Coral Sea St Ne, Mounds View, Minnesota, United States
FEI 2182208
FDA has 14 inspections on file for Medtronic, Inc. in Mounds View, Minnesota (FEI 2182208), from 2009 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below. One of the three FOIA requests for this site has asked FDA for this inspection's 483.
The most recent OAI inspection ended on May 14, 2018. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.100(b) and 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 14, 2009–2025
- 2 OAI · 5 VAI · 7 NAI
- Inspections with 483 observations shown
- 4 of 14
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Mar 20, 2025
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
14 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch MonitoringCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a process have not been adequately established.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of procedures, or not maintained
Document control procedures were not implemented.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Procedures
The procedures addressing documentation of corrective and preventive action activities were not implemented.
Design changes - Lack of or Inadequate Procedures
Procedures were not followed for the validation or verification of design changes before their implementation.
Show 3 more citations
Investigation procedures
The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not implemented.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Evaluation of suppliers, contractors, etc., requirements
Adequate quality requirements that must be met by suppliers were not implemented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Mar 20, 2025.
Inspection ending Feb 10, 2025VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Mar 20, 2025FDA-FOIA-2025-2575Form 483· 1 of 4 sitesClosed
Inspection ending Feb 4, 2020VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Nov 8, 2023FDA-FOIA-2023-9940Form 483No records found
Inspection ending May 14, 2018OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Feb 23, 2021FDA-FOIA-2021-1216EIRDelogged (not a FOIA Request)
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 14 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action4 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.50Purchasing controls2 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.70Production and process controls1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Other FEIs recorded as Medtronic, Inc.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Medtronic, Inc.FEI 3015544746Minneapolis, Minnesota, United States
Last inspected 2025-12-10VAI
Medtronic, Inc.FEI 3015180993Sunnyvale, California, United States
Last inspected 2020-10-08NAI
Medtronic, Inc.FEI 3004612291Bellevue, Washington, United States
Last inspected 2013-01-24NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2182208 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2182208. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.