FDA has one inspection on file for BioDiagnostic International in Brea, California (FEI 2023751). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDER).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 10 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below.

No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2018
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Cleaning / Sanitizing / Maintenance

      Equipment and utensils are not cleaned, maintained and sanitized at appropriate intervals to prevent contamination that would alter the safety, identity, strength, quality or purity of the drug product.

    2. Storage under appropriate conditions

      Drug products are not stored under appropriate conditions of temperature, humidity and light so that their identity, strength, quality, and purity are not affected.

    3. Testing and release for distribution

      Testing and release of drug product for distribution do not include appropriate laboratory determination of satisfactory conformance to the final specifications and identity and strength of each active ingredient prior to release.

    4. Adequate space lacking to prevent mix-ups and contamination

      The building lacks adequate space for the orderly placement of equipment and materials to prevent mix-ups between drug product containers, closures, in-process materials and drug products and to prevent contamination.

    5. Testing Each Component for Conformity with Specs

      Each component is not tested for conformity with all appropriate written specifications for purity, strength, and quality.

    Show 5 more citations
    1. Lack of quality control unit

      There is no quality control unit.

    2. Absence of Written Procedures

      There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess.

    3. Lack of written stability program

      There is no written testing program designed to assess the stability characteristics of drug products.

    4. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate specifications designed to assure that components and drug products conform to appropriate standards of identity, strength, quality and purity.

    5. Equipment Design, Size and Location

      Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design, of adequate size and suitably located to facilitate operations for its intended use and cleaning and maintenance.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.

Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDER · DrugsFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2023751 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2023751. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.